CALIBRATION: ZERO SET

The refractometer must be set to zero before initial use and periodically 

thereafter. It is recommended that calibration be performed at least 

once a day as well as prior to performing tests requiring the highest 

precision, or when moving between environments with extreme 

changes in ambient temperature. A clean container of water is all that 

is needed to automatically calibrate the instrument. Although tap water 

may be used, distilled or deionized water is recommended. The water 

temperature for calibration should ideally near 68°F (20°C). 

1.  Inspect the measuring surface to make sure  

it is clean and dry.

2.  Place a few drops of calibration liquids on the  

prism window, making sure to eliminate any  

bubbles.

3. Press   to start calibration.

4.  The unit shows the process of zero setting  

(Fig 2.1).

5.  If calibration is successful, it will show “Pass”  

(Fig 2.2).

6. If calibration is not successful, it will show “Fail”.

7.  After calibration process is finished, the LCD will 

return to temperature mode. The calibration result 

will be saved and will be the new zero point after  

the device is powered off and on again.

Note: After calibration, please remember to clean the liquid from  

the prism surface and keep the prism surface dry and clean to  

avoid erosion.

 

MEASUREMENT

1. Make sure the prism surface is clean and dry.

2. Place a few drops of the sample on the prism.

3. Press   key, “---” will be shown on the screen. 

–  Test result(s) will be shown on the screen  

(Fig 3.1).

–  The result(s) will be stored in the device  

for up to 60 seconds. 

– Press the “Power” button to recheck the previous measurements.

4.  After testing, please remember to clean and dry the surface when 

you are done.

Scale Selection

1.   Hold 

 for 2 seconds, the scale will change to  

the next measurement type (Fig 4.1).

2.  Repeat Step 1 until desired scale is shown  

on the screen.

3. The device saves the last selected scale.

Temperature Selection

1.   Hold 

 for 2 seconds, the temperature unit will change between  

°F and °C.

2. The device saves the last selected temperature unit.

Resetting

1.   Press 

 +  , the option of returning to factory  

setting of zero will be shown on the screen  

(Fig 5.1).

2.   Press 

 button to confirm resetting, or  

press   to cancel.

Fig 2.1

Fig 2.2

Fig 3.1

STATUS

1. Press   to turn on the device.

2.   Press 

 to turn off the device; the LCD will  

show the process (Fig 6.1).

3.  To conserve energy, the device enters Sleep  

Mode after standing by for more than 60 seconds. 

Press   to wake up the device.

4. The refractometer will turn off after 90 seconds of nonuse. 

TROUBLESHOOTING

Zero setting error (Fig 7.1)

– Out of calibration (zero set) range.

Action needed: Make sure there is a sufficient 

amount of clean calibration water to cover the  

prism surface.

Measurement Errors

–  High – exceeds the specified measurement range 

(Fig 7.2).

–  Low – falls below the specified measurement range 

(Fig 7.3).

–  Error – no liquid on the prism or not sufficient 

amount of liquid (Fig 7.4). 

Temperature errors

–  High – temperature exceeds range (Fig 7.5).

Action needed: Check temperatures of the sample 

liquid, the device, and the ambient surroundings.

–  Low – temperature falls under range (Fig 7.6).

Action needed: Check temperatures of the sample 

liquid, the device, and the ambient surroundings.

Battery Error

–  Low – battery indication (Fig 7.7).

Action needed: Replace with two new AAA batteries 

when the battery icon is red.