LCD DISPLAY

KEYS (BUTTONS)

1. Turn on meter; or

2. Turn off meter after pressing for 3 seconds.

1. Zero calibration; or

2. Press for 2 seconds to toggle between °C and °F.
1. Press the button to read after drop samples; or

2.  Press for 2 seconds to toggle between measured 

items.

 

+

 

Reset to factory setting.

INTRODUCTION

Thank you for choosing our refractometer. This is an easy-to-use 

device, requiring little to no training. Please read the manual before 

using to ensure optimal measurement. 

The refractometer is designed for fast and accurate determination of 

fluid concentrations. The instrument will automatically compensate  

for temperature fluctuation and can be used for years of trouble-free 

service when properly and regularly maintained. Instrument has  

an IP65 waterproof rating.

The refractometer represents reading relative to Brix scale. The Brix 

scale originated in the food industry and is primarily a unit of measure 

corresponding to percent of sugar in a sugar and water solution. The 

actual Brix value represents the number of grams of cane sugar in a 

100-g cane-sugar solution (percent sugar wt./wt.). This direct reading 

relationship holds true only for sucrose solutions. When measuring 

non-sugar solutions, the Brix scale should be thought of as an arbitrary 

scale. To make these arbitrary units of measure meaningful, they must 

somehow be correlated to the concentration on the solution you are 

testing. This is accomplished by creating a chart of solution concentra-

tion relative to the Brix scale. A separate must be made to each type of 

solution being tested.

Unpacking

Check individual parts against the list of items below. If anything  

is missing or damaged, please contact your instrument supplier  

immediately.

1. Refractometer

2. Soft and clean cloth

3. Hard carrying case

4. Operation manual

5. Two AAA batteries

SPECIFICATIONS

Temperature measurement range: 32 to 104°F (0 to 40°C)

Measurement range: Please see table above. 

Precision of measurement temperature: ±1°F (0.5°C)

Automatic temperature compensation: 41 to 104°F (5 to 40°C) 

Minimum sample volume: 0.3 mL 

Measuring response time: 

3 seconds 

Power supply: two AAA batteries 

Battery life: 

5000 readings 

Dimensions (L x W x H): 5.7" x 2.6" x 1.5"(145 × 67 × 38 mm) 

Weight: 6.5 oz (185 g)

AUTO TEMPERATURE COMPENSATION (ATC)

Refractive index (RI) is very temperature-dependent. It is well known 

that substantially all materials expand when heated (i.e. become less 

dense) and contract when cooled (i.e. become more dense). The speed 

of light in a liquid increases as temperature increases, and refractive 

index, therefore, decreases. Although this thermal effect is minor for 

solids, the change in density for a liquid is substantial. ATC ensures  

that concentration readings of aqueous (water-based) solutions will  

be accurate with respect to sample's temperature.

The refractometer is temperature compensated for aqueous (water-

based) sucrose solutions and can automatically compensate for  

temperature differences within the range of 41 to 104°F (5 to 40°C). The 

temperature of the sample, however, has little bearing on the accuracy 

of the reading. In most cases, the sample almost immediately assumes 

the temperature of the refractometer, the ambient temperature, and the 

fluid should be in equilibrium within the instrument's temperature 

range. 

Range

Resolution

Accuracy

ATC

Catalog  

number 

0 to 45% Brix, 1.3330 to 1.4098 RI

0.1, 0.0001  ±0.2, ±0.0003  Yes

81150-48

0 to 85% Brix, 1.3330 to 1.5100 RI

0.1, 0.0001  ±0.5, ±0.0003  Yes

81150-55

0 to 95% Brix, 1.3330 to 1.5400 RI

0.1, 0.0001  ±0.5, ±0.0003  Yes

81150-56

0 to 28% NaCl, 1.3330 to 1.4100 RI

0.1, 0.0001  ±0.2, ±0.0003  Yes

81150-49

0 to 45% Brix, 0 to 28% NaCl, 

1.3330 to 1.4100 RI

0.1, 0.1,  

0.0001 

±0.2, ±0.2, 

±0.0003 

Yes

81150-57

0 to 60% v/v, w/w ethyl alcohol

0.1

±0.3

Yes

81150-52

0 to 60% v/v, w/w, isopropyl 

alcohol; –58 to 32°F, –50 to 0°C

0.1

±0.5

Yes

81150-53

0 to 75% ethylene glycol,  

–58 to 32°F, –50 to 0°C

0.1

±0.3, ±1°F, 

±0.5°C

Yes

81150-54