11.2 Supplementary Manual for Metallurgical Model 78904-30 (cont.)

Components 

A. 

Microscope head

B. 

Aperture diaphragm 

C. 

Field diaphragm

D. 

Two slots for color filters 

E. 

Condenser focusing knob

F. 

Brightness adjustment knob

G. 

Main body of metallurgical attachment

H. 

Power adapter input

1

22

Содержание 78904-16

Страница 1: ...Installation and Operating Instructions MSU 600 Series Compound Microscopes Models 78904 16 18 21 24 27 30 33...

Страница 2: ...____________________________________________________________________ 13 5 12 Auto Off Function ______________________________________________________________________ 13 6 0 Phase contrast model 78904...

Страница 3: ...enclosed with a cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated...

Страница 4: ...or non alcohol based disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer a RMA return authorization form and a decontamination...

Страница 5: ...se undue force when turning the knobs Make sure that the microscope system can dissipate its heat fire hazard Keep the microscope approximately 15 cm free from walls and obstructions Never turn the mi...

Страница 6: ...04 16 Binocular 4x 10x S40x S100x semi plan 48 to 76 mm Rackless integrated X Y mechanical stage 3 W NeoLED 100 240 VAC 50 60 Hz 24 1 lb 10 9 kg 78904 18 Trinocular 4x 10x S40x S100x semi plan 48 to 7...

Страница 7: ...e J Nosepiece for 5 objectives C Microscope head K Objectives D Height adjustment condenser L Stage with X Y mechanical stage E Slide protection handle M Condenser with iris diaphragm F Coaxial coarse...

Страница 8: ...sure the power switch is turned off the plug removed before replacing the led unit or fuse 4 1 Assembling steps The steps mentioned below are often not necessary but described for your convenience no...

Страница 9: ...l Each eyepiece has a rubber eyeshade This prevents damage to the lens and stray light The eyeshade can simply be slipped over the eyepiece Connecting the power cord The MSU 600 series supports a wide...

Страница 10: ...of the specimen holder backwards Release the arm slowly clamping the slide with the cover glass facing up Rotating the X and Y axis knob will move the specimen to the center for alignment with the ce...

Страница 11: ...one diopter adjustment Position the diopter marking on the zero point Look into both eyepieces and focus on the specimen Close the right eye and look into the eyepiece with diopter adjustment rotate...

Страница 12: ...ion is obtained When contrast is low rotate the diaphragm adjustment ring to 70 to 80 of the N A of objective This will improve the contrast of the image The diaphragm is factory pre centered 5 9 Use...

Страница 13: ...image appears in the eyepiece 6 2 Using the Phase contrast slider Keep the phase contrast slider face up text up insert it from left to right into the condenser slider socket as the direction of the...

Страница 14: ...w If the ring images can t be observed clearly first try and focus the centering telescope If this does not solve the issue raise or decline the condenser If the bright ring and dark ring images are n...

Страница 15: ...blow in order to prevent internal damage to the microscope And in most cases replacing with a correctly rated fuse will resolve the issue However should you encounter a blown fuse frequently please co...

Страница 16: ...orm the height adjustment procedure above while watching the image on the computer screen In this case once you have obtained parfocality in the device tighten screw A again Screw B is only used to fi...

Страница 17: ...___________________________________________________ ___________________________________________________________________________________________________ ________________________________________________...

Страница 18: ...11 0 Supplementary Manuals 11 1 Supplementary Manual for Darkfield Model 78904 27 18...

Страница 19: ...ndenser by unwinding the screw marked with the arrow Set up the dark field condenser and tighten the screw 3 Lower the condenser slightly place the slide In this way the immersion oil should not yet t...

Страница 20: ...is aperture you must also put immersion oil on the prepared slided 9 Focus the lens On the S100x 1 25 oil immersion objective A and S100x 1 25 IOS super contrast objective B you can adjust the backgro...

Страница 21: ...11 2 Supplementary Manual for Metallurgical Model 78904 30 21...

Страница 22: ...l Model 78904 30 cont Components A Microscope head B Aperture diaphragm C Field diaphragm D Two slots for color filters E Condenser focusing knob F Brightness adjustment knob G Main body of metallurgi...

Страница 23: ...ing the Illumination 1 Adjust the condenser knob 1 E to the position where the image brightness in view is homogenous 2 Adjust the field diaphragm 1 C and aperture diaphragm 2 B until the two diaphrag...

Страница 24: ...11 3 Supplementary Manual for Polarization Model 78904 33 24...

Страница 25: ...e rotation clamping knob 1 is loosened the stage can be rotated 360 horizontally Using the Bertrand Lens By revolving the Bertrand lens dial 3 the Bertrand lens can be selected At the O position the l...

Страница 26: ...de approximately half way to the eyepiece crosshairs c 9 Rotate the stage back to its 0 position 10 The 4x objective is equipped with two adjustable screws inside the revolving nosepiece for centering...

Страница 27: ...n be rotated 360 degrees turn it until complete extinction is obtained 5 Place the specimen for orthoscopic observation 6 Insert test plates for further observation test and study Conoscopic observati...

Страница 28: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 600_Series_Compound Microscopes_Manual...