52412Mechnicaloven_ver_1.2
53
4.4.4
Alarm
It is a function to generate an alarm when the temperature is above or below the user-specified
temperature or when the door is opened [4.5.3 Refers to Alarm Display and Pop-up Notification
Function].
Item
Function
Input Value
Default
High Deviation
Temperature
Set the upper limit of temperature
deviation.
If the actual temperature becomes
higher than the target tempe
allowable upper limit temperature, an
alarm is generated.
0 ~ 250
℃
5
Low
Deviation
Temperature
Set the lower limit of temperature
deviation.
If the actual temperature becomes
lower than the target temperature -
allowable lower limit temperature, an
alarm is generated.
0 ~ 250
℃
5
High
Temperature
Set the absolute upper limit
temperature.
An alarm occurs if the actual
temperature exceeds the absolute
upper limit temperature.
0 ~ 270
℃
270
Door Open
Alarm Delay
Set the alarm for the door open delay
time.
If the door opens exceeding Door
Open Alarm Delay, an alarm occurs.
Off (0) ~ 30min
5
Содержание 52412-15
Страница 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...
Страница 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...
Страница 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...
Страница 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...
Страница 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...
Страница 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...
Страница 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...
Страница 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...
Страница 69: ...52412Mechnicaloven_ver_1 2 69 8 0Appendix...