52412Mechnicaloven_ver_1.2
21
(7)
Side panel-2
D) Main switch
It turns on and off the main power.
E) Communication port (Option)
The USB port and the RS232 port connects the device to a computer to monitor or operate the
instrument. When connecting both ports at the same time, the USB will be recognized firstly.
If communication port is not selected as an option,
attached and the product is
manufactured.
F) Over Temperature Limiter
If the heat rises excessive temperature, the temperature control stops, and a warning sign will be
displayed on the screen along with a notification sound. Once you check the warning sign, please
solve the problem, and reset the Over Temperature Limiter up to 15% higher than the original
target temperature. You can restart the instrument by pressing the Start button.
G) Product Label
The label describes the instrument name and its type.
(8)
Foot
It supports its body and prevents slipping.
G
F
E
D
Содержание 52412-15
Страница 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...
Страница 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...
Страница 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...
Страница 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...
Страница 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...
Страница 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...
Страница 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...
Страница 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...
Страница 69: ...52412Mechnicaloven_ver_1 2 69 8 0Appendix...