Cole Parmer 52412-15 Скачать руководство пользователя страница 20

 

52412Mechnicaloven_ver_1.2 

                                                                                                                 20 

 

 

(4) 

Ventilating hole 

To control the amount of air to be exhausted, you can use the ventilating hole.   

 

(5) 

Door handle 
The door handle can open or close the door.   

(6) 

Side panel-1 

 

 

A)  BMS connection port (Not selectable) 

The BMS function is not provided in this model. 

 

B) Circuit breaker 

A circuit breaker is embedded to protect the device from short circuits or overcurrent. 

 

C) Power cord socket. 

The socket connects the power line to the instrument.     

 

 

 

 

 

• The instrument is in excessive heat, therefore please were gloves when controlling the 
exhaust. Wet gloves can cause burns on the users’ hands, since it speeds up the 
temperature conduction.   
• Also, wet gloves or bare hands are not allowed, since the operator can be electrocuted 
when a short circuit occurs due to aging of the device. 

A

 

C

 

B

 

Содержание 52412-15

Страница 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...

Страница 2: ...t Disclaimer The material in this manual is for the users information only The products it describes are subject to change without prior notice due to the manufacturer s continuous development program...

Страница 3: ...reading 11 1 4 1 Warnings 12 1 4 2 Caution 14 2 0 Functional Description 16 2 1 Introduction 17 2 2 Feature 17 2 2 1 Excellent Performance 17 2 2 2 Safety Enhancement 18 2 2 3 Improved Convenience 18...

Страница 4: ...uto Tuning 51 4 4 4 Alarm 53 4 4 5 Password 54 4 4 6 Factory Default 54 4 5 Safety Function 55 4 5 1 Safety Functions Related to Temperature 55 4 5 2 Safety Functions Related to Use 55 4 5 3 Alarm Dis...

Страница 5: ...of Accessories 68 8 0 Appendix 69 8 1 Technical Specification 70 8 1 1 OVF 800 Series 70 8 2 Disposal of Product 71 8 3 Warranty 71 8 3 1 Warranty period 71 8 3 2 Warranty Exception 71 8 3 3 How to Re...

Страница 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...

Страница 7: ...atures and structure of the instrument The chapter 3 contains installation instructions for instrument confirmation and installation Also it describes preparation procedures before operation and insta...

Страница 8: ...ciency and productivity is to establish effective safety programs and comply with safety rules In addition to maintaining and repairing the instrument regularly please check the condition of the instr...

Страница 9: ...the instrument If you need to move the instrument or use it after a long pause make sure that the instrument is operating normally Please stop using the instrument when it appears with the following...

Страница 10: ...nts Indicates a hazardous situation which if not avoided will result in death fatal injury or property damage Indicates a hazardous situation which if not avoided could result in serious injury or pro...

Страница 11: ...fety Symbols The graphics below are used to convey important information about the use and characteristics of the instrument without further explanation Non safety Symbols Permissible ambient temperat...

Страница 12: ...ower connections can cause damage to the instrument and serious personal injury or death Do not handle or touch the electrical cords or electrical components with wet hands 3 Installation Avoid contac...

Страница 13: ...nge sounds smells or smoke Do NOT operate damaged or short circuit equipment Do NOT move the instrument when the power cord is connected or when the instrument is operating 5 Repair and Modification D...

Страница 14: ...chine Operation When initiating the instrument the operator should check the machine continuously until it reaches to normal status Wear gloves when handling samples to prevent burns since there may b...

Страница 15: ...its surroundings Do not disassemble when cleaning the instrument If you are cleaning the instrument in a way other than the above please check the cleaning manner with Cole Parmer 6 Repair The electri...

Страница 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...

Страница 17: ...rs the heat to the samples quickly evenly and precisely 3 A reliability and reproducibility of temperature An excellent reliability and reproducibility of temperature distribution has been verified by...

Страница 18: ...een and the alarm sound is activated 6 Instantaneous Outage Recovery Features If the instrument has a short period of power outage occurs within 30 seconds the system returns to the previous operating...

Страница 19: ...the vent hole cover 3 Controller Various controllers and electrical components are embedded in the instrument Since the instrument is in excessive heat please wear gloves when controlling the vent ho...

Страница 20: ...aker A circuit breaker is embedded to protect the device from short circuits or overcurrent C Power cord socket The socket connects the power line to the instrument The instrument is in excessive heat...

Страница 21: ...ttached and the product is manufactured F Over Temperature Limiter If the heat rises excessive temperature the temperature control stops and a warning sign will be displayed on the screen along with a...

Страница 22: ...he heater If the door is opened for more than 2 minutes the displays will show you an error message with an alarm sound Moreover the instrument operation will be stopped with the power disconnection w...

Страница 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...

Страница 24: ...to the instrument and make sure it matches the model you ordered If there is any component below that does not match with your order please contact the branch or device seller of Cole Parmer Please r...

Страница 25: ...nstall the instrument in a place with a possibility of flammable gas leakage Do not use or store flammable dust combustible gas around this instrument Operation of the device is prohibited in a place...

Страница 26: ...rument on a flat and solid floor for any case of an accident Since the door is opened to the left with maximum 115 please consider sufficient installation space Please keep the appropriate space typic...

Страница 27: ...wn in the figure below It is also recommended to install it in an environment equipped with ventilation facilities such as a fume hood or arm hood unit mm Please avoid samples or shelves contacted to...

Страница 28: ...he height of the shelf by attaching the shelf support to the shelf adjustor on the side of the chamber as shown below Place the shelf on the shelf support attached to the shelf adjustor Depending on t...

Страница 29: ...55 710 C Interior depth 210 330 350 376 D Shelf width 266 346 451 516 E Shelf depth 195 315 335 361 F The height between the top hole of the shelf installation and the inner top 80 90 87 5 85 G Height...

Страница 30: ...nformation about potential risks or how to avoid them For safe operation all users must check the safety label before using it and read the 1 4 Instructions for Safe Use required reading carefully to...

Страница 31: ...put terminal on the side of the body 3 Connect the other side of the power plug to a power outlet 4 Turn on the power switch and circuit breaker to supply power to the device Before installing the pro...

Страница 32: ...n This device is not an explosion proof structure Do not place explosive flammable flammable or oxidative chemicals such as alcohol benzene gasoline on top of the device store them or put them in the...

Страница 33: ...there may be residual heat even after the power is off Before using the instrument please make sure the inside of the chamber is clean all times If this instrument has moisture please stop operation i...

Страница 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...

Страница 35: ...ervation end reservation setting temperature control and stop 4 2 2 General Settings You can check and change the settings of the device such as sound settings date time modification and language sele...

Страница 36: ...tton to make general and system settings 2 You can set the target temperature value 3 Warnings When a warning occurs an alarm sign appears and a pop up notification window appears 4 Current Temperatur...

Страница 37: ...ot operate with a door open sound Please close the door and restart it If the door is opened while the device is operating the fan and heater will stop operating and a door open warning message will b...

Страница 38: ...on without pressing Enter the temperature settings will be canceled The set temperature can be input from 0 to 250 and the temperature control range of the device is Amb 10 250 To set and start the te...

Страница 39: ...l of the device after a set time And after reaching the set temperature there is a wait off timer function to control the temperature maintenance for a set time 1 Move to Timer Settings by pressing th...

Страница 40: ...imer Set with the unit of hh mm In case setting the start timer as 1 hours 4 You can return to the Basic Mode screen by pressing button with the Wait ON OFF timer applied When the timer setting is com...

Страница 41: ...Timer is ON Wait OFF Timer is ON If you only set the Wait ON Timer as ON the instrument waits for the input value before raising the temperature as the set value In this case the heater will be contin...

Страница 42: ...etting temperature If you set up the Wait OFF Timer the Wait OFF Timer works automatically once the set temperature has reached In case Wait OFF Timer is set 1 If you press the start button after sett...

Страница 43: ...on again If you want to change the Wait OFF timer during timer operation you must turn off the Wait OFF Timer and reset the process 4 2 5 Instantaneous Outage Recovery Function If the power is restore...

Страница 44: ...Sound Display Clock Language Auto Run Safety Lock Door Event Communication Pressing the General Settings button in the home screen takes you to the General Settings 4 3 1 Sound Display In Sound Displ...

Страница 45: ...n 1 OVF 800 Series does not support the on lamp by door 4 3 2 Clock The Clock is a function that sets the time within the device The time set in the Clock is the standard for the date and time of the...

Страница 46: ...orted later 4 3 4 Event Event is the function to indicate the start of control and errors occurred in the device The latest event occurrence and time are displayed Up to 4 events can be displayed on o...

Страница 47: ...on It is the function that allows communication settings with LC GreenBox Option Item Function Value Range Initial Value Equipment Address Set communication Address 1 255 1 Baud Rate Set Communication...

Страница 48: ...he user sets the Auto Run and the power is restored after the device is turned off for 30 seconds or more during temperature control follow the procedure below Previous States Operation in Basic Contr...

Страница 49: ...Temperature Input Calibration Auto Tuning Alarm Password and Factory Default 1 If you press the System Settings button on home screen the password window appears 2 After entering the set password pre...

Страница 50: ...e decimal place 0 1 1 1 4 4 2 Calibration When measuring the temperature inside the chamber by using the user s temperature sensor if there is a temperature difference between the temperature sensor i...

Страница 51: ...etup Calibration and enter Temperature 1 Temperature 2 and Temperature 3 in the Uncalibrated column Also enter the internal temperature measured by the external temperature sensor in the Calibrated co...

Страница 52: ...in program control 3 It moves directly to the operation display window of Basic Mode and the icon on the top status bar blinks indicating that Auto Tuning is running 4 When Auto Tuning is completed t...

Страница 53: ...the target temperature allowable upper limit temperature an alarm is generated 0 250 5 Low Deviation Temperature Set the lower limit of temperature deviation If the actual temperature becomes lower t...

Страница 54: ...nce If you press the reset button a pop up notification appears and confirms whether or not to proceed with Factory Default This function initializes all settings changed in the Basic Settings General...

Страница 55: ...sage for door open appears But there is no alarm sound If the door is closed within the Door Open Alarm Delay time default setting 5 minutes the previous operation continues If the door is not closed...

Страница 56: ...is higher than the target temperature allowable upper limit temperature Alarm displays when the actual temperature is lower than the target temperature allowable lower limit temperature Alarm displays...

Страница 57: ...uld be stabilized for a certain period of time 3 Turn the red knob counterclockwise using a screwdriver Stop the knob when a warning is displayed on the screen with a notification sound 4 Turn the red...

Страница 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...

Страница 59: ...value Check function of controller Electric wire related checklist Defects if electric wire Check contamination of electric wire Check fixation of wiring connection part Before using the instrument pl...

Страница 60: ...t 1 In general Step 1 Take out the shelf from the instrument and wipe it with a dry cloth using a neutral detergent Step2 Wipe the inner surface of the instrument with a dry cloth using a neutral dete...

Страница 61: ...h or a sponge to wash the cleaning tool with a neutral detergent Do not pour water directly into the instrument while cleaning Do not arbitrarily disassemble and clean the instrument If cleaning the i...

Страница 62: ...rtation 6 When the instrument is not in use for a long period of time unplug the power plug clean it pack it and store it in a dry place Do not move the instrument with the power cord connected Do not...

Страница 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...

Страница 64: ...amaged outlet plug power cord Power supply does not match the equipment s power type Check the voltage phase and capacity of the power supply and connect the power correctly Either circuit breaker is...

Страница 65: ...vice failure Request service In case the temperature control is not working Please check whether there is any device that generates strong high frequency noise around the instrument Please remove such...

Страница 66: ...the instrument and restart it after the instrument has cooled down If the same problem occurs request service door is open Check if the door text on the display window is on and close the door securel...

Страница 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...

Страница 68: ...F 800 26 52412 31 Wire Shelf for OVF 800 61 52412 32 Wire Shelf for OVF 800 112 52412 33 Wire Shelf for OVF 800 147 52412 34 Perforated Shelf for OVF 800 26 52412 35 Perforated Shelf for OVF 800 61 52...

Страница 69: ...52412Mechnicaloven_ver_1 2 69 8 0Appendix...

Страница 70: ...eel wire electro polished Insulation Mineral wool Ventilation slide Stainless steel dia 38mm x 1ea Dimensions Interior W x D x H mm inch 300x210x400 11 8x8 27x15 75 380x330x480 14 97x13 0x18 9 485x350...

Страница 71: ...ratory equipment and its components 8 3 Warranty 8 3 1 Warranty period 1 If the product fails during normal and proper use the warranty period for manufacturing liability shall be two years from the d...

Страница 72: ...Details of the Product If you would like to purchase any product accessories please contact the following branch office or head office 8 3 4 Returning Process of the Product If you need to return the...

Страница 73: ...E de sales antylia com W coleparmer de T 1 847 549 7600 E sales antylia com W coleparmer com France USA T 33 0 1486 37800 E fr sales antylia com W coleparmer fr T 1 847 549 7600 E sales antylia com W...

Отзывы: