Cole Parmer 52411-00 Скачать руководство пользователя страница 9

                                                                                                   

52411Dryingoven_ver_1.0

 

www.coleparmer.com

                                                                                                                                                                       

 

Natural Convection Drying Oven 

 

Model 

52411-00 

52411-02 

52411-04 

52411-06 

52411-08 

52411-01 

52411-03 

52411-05 

52411-07 

Electrical Requirement 

120V 60Hz,220-240V 50-60Hz 

220-240V 

50/60Hz 

Temperature range 

RT+10

℃~

200

 

Temperature stability 

±

0.5°C

 

Power consumed 

850W 

1000W 

1400W 

2000W 

2200W 

Cable Replacement 

120V 

– 16AWG 

220V 

– 1mm

2

/18AWG 

120V 

– 14AWG 

220V 

– 1mm

2

/18AWG 

120V 

– 12AWG 

220V 

– 1mm

2

/18AWG

 

120V 

– 12AWG 

220V 

– 1.5mm

2

/14AWG

 

1.5mm

2

/14AWG

 

Volume 

27L 

56L 

96L 

140L 

200L 

Fuse / Circuit breaker 

120V-10A Circuit Breaker, 

220V - 5A, 5 x 20 mm, glass 

fast blow 

120V-15A, 6 x 32 mm, 

ceramic fast blow, 

220V - 7A, 5 x 20 mm, 

glass fast blow 

120V-20A, 6 x 32 mm, 

ceramic fast blow, 

220V - 10A, 5 x 20 

mm, glass fast blow 

120V-20A Circuit 

Breaker, 

220V - 15A, 6 x 32 mm, 

glass fast blow 

16A Circuit breaker 

Interior Dimensions   

(W x D x H) 

320x300x355mm 

400x330x415mm 

450x430x505mm 

520x500x575mm 

570x560x640mm 

 
 

Temperature Controller Detailed Operation

 

5.1 Switch on   

1

Connect the power according to the power outlet with a protective ground connection.   

2.  Turn on  the  instrument by pressing the  ON/OFF  Button and the  screen  will  light up.  The 

display shows the initialization sequence before the instrument is ready to use.   

NOTE: Every time you turn on the instrument it beeps intermittently, the icon of visual alarm 

and the word 

“end” will appear on the display, indicating that a heating cycle had been done 

before. Press any button to silence the audible signal and the icon appears. 

 

5.2 Setting of parameters 

5.2.1 Time Settings 

1

After confirming the temperature, the last value of the set time (timer) will start to flash. Set 

the desired value (hh:mm) by pressing 

or

 

keys. It’s possible for a quick movement 

between the digits by using 

button. 

2

Confirm the set value with another press of 

 

button.   

Содержание 52411-00

Страница 1: ...52411Dryingoven_ver_1 0 Forced Air Oven Natural Convection Oven User Manual...

Страница 2: ...sed by failure to follow instructions lack of maintenance and any unauthorized modification 2 Contents of package The instrument is delivered complete with the following parts Drying Oven main unit 2...

Страница 3: ...for connecting to the mains power supply replace the plug with a suitable alternative THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN NOTE Refer to the equipment rating plate to en...

Страница 4: ...ng surface when the set temperature is over 176 F 80 C 8 Do not put hands or objects into the air inlet or air outlet 9 Unauthorized repair is not allowed The unit should have routine inspections and...

Страница 5: ...placed on solid surface to keep it horizontal 2 Aggressively opening or closing the door may cause damage to the door oven or injury to users 4 Product introductions 4 1 Function introduction Control...

Страница 6: ...52411Dryingoven_ver_1 0 www coleparmer com 4 2 Display and commands 4 2 1 Display introduction Running status Temperature Set Temperature Fan Status Buzzer MUTE ALARM Timer...

Страница 7: ...change the digit Program units tens etc of the value of the parameter you are editing 3 In combination with the key allows access to menus with password 4 and adjustment buttons allow users to increas...

Страница 8: ...aker 220V 7A 5 x 20 mm glass fast blow 120V 16A Circuit Breaker 220V 10A 5 x 20 mm glass fast blow Interior Dimensions W x D x H 250x260x250mm 340x320x320mm 420x395x350mm 450x400x450mm Model 52411 17...

Страница 9: ...uit breaker Interior Dimensions W x D x H 320x300x355mm 400x330x415mm 450x430x505mm 520x500x575mm 570x560x640mm 5 Temperature Controller Detailed Operation 5 1 Switch on 1 Connect the power according...

Страница 10: ...tion 1 After setting the operating parameters press button for about 4 5 seconds to start the heating cycle cooling process with defined time in hh mm or continue the process for 00 00 When the word E...

Страница 11: ...arts the work cycle but it doesn t immediately heat the word end and the set delay time alternately blink on the top right part of display counting the wait time until the real starting of heating Onc...

Страница 12: ...s Temperature offset on single point After set the Pb value Press to exit 6 2 Other parameters function in Menu Here are the passwords and access sequence to various parameters functions PASSWORD FUNC...

Страница 13: ...he instrument must be returned for service it is necessary to provide for proper cleaning and possible decontamination for pathogens The original package is recommended for the product if it needs to...

Страница 14: ...ing Abnormal temperature display Sensor is out or wiring is broken Repair or replace Pt100 No temperature rise Check if timing is set up and time is up Refer to the operation of timing function The co...