![Cole Parmer 52411-00 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/cole-parmer/52411-00/52411-00_user-manual_2635448009.webp)
52411Dryingoven_ver_1.0
Natural Convection Drying Oven
Model
52411-00
52411-02
52411-04
52411-06
52411-08
52411-01
52411-03
52411-05
52411-07
Electrical Requirement
120V 60Hz,220-240V 50-60Hz
220-240V
50/60Hz
Temperature range
RT+10
℃~
200
℃
Temperature stability
±
0.5°C
Power consumed
850W
1000W
1400W
2000W
2200W
Cable Replacement
120V
– 16AWG
220V
– 1mm
2
/18AWG
120V
– 14AWG
220V
– 1mm
2
/18AWG
120V
– 12AWG
220V
– 1mm
2
/18AWG
120V
– 12AWG
220V
– 1.5mm
2
/14AWG
1.5mm
2
/14AWG
Volume
27L
56L
96L
140L
200L
Fuse / Circuit breaker
120V-10A Circuit Breaker,
220V - 5A, 5 x 20 mm, glass
fast blow
120V-15A, 6 x 32 mm,
ceramic fast blow,
220V - 7A, 5 x 20 mm,
glass fast blow
120V-20A, 6 x 32 mm,
ceramic fast blow,
220V - 10A, 5 x 20
mm, glass fast blow
120V-20A Circuit
Breaker,
220V - 15A, 6 x 32 mm,
glass fast blow
16A Circuit breaker
Interior Dimensions
(W x D x H)
320x300x355mm
400x330x415mm
450x430x505mm
520x500x575mm
570x560x640mm
5
Temperature Controller Detailed Operation
5.1 Switch on
1
.
Connect the power according to the power outlet with a protective ground connection.
2. Turn on the instrument by pressing the ON/OFF Button and the screen will light up. The
display shows the initialization sequence before the instrument is ready to use.
NOTE: Every time you turn on the instrument it beeps intermittently, the icon of visual alarm
and the word
“end” will appear on the display, indicating that a heating cycle had been done
before. Press any button to silence the audible signal and the icon appears.
5.2 Setting of parameters
5.2.1 Time Settings
1
.
After confirming the temperature, the last value of the set time (timer) will start to flash. Set
the desired value (hh:mm) by pressing
or
keys. It’s possible for a quick movement
between the digits by using
button.
2
.
Confirm the set value with another press of
button.