4.3

 

Specifications 

Forced Air Drying Oven 
 

Model 

52411-09 

52411-11 

52411-13 

52411-15 

52411-10 

52411-12 

52411-14 

52411-16 

Electrical Requirement 

120V 60Hz,220-240V 50-60Hz 

Temperature range 

RT+10~250

 

Temperature fluctuation 



 

Power consumed 

600W 

850W 

1100W 

1550W 

Cable Replacement 

120V 

– 16AWG,   

220V 

– 1mm

2

/18AWG 

120V 

– 14AWG,   

220V - 1mm

2

/18AWG 

Volume 

16L 

30L 

50L 

80L 

Fuse / Circuit breaker 

120V-10A Circuit Breaker, 

220V - 5A, 5 x 20 mm, glass fast blow 

120V-10A Circuit Breaker, 

220V - 7A, 5 x 20 mm, glass 

fast blow 

120V-16A Circuit Breaker, 

220V - 10A, 5 x 20 mm, glass 

fast blow 

Interior Dimensions (W x D x H) 

250x260x250mm 

340x320x320mm 

420x395x350mm 

450x400x450mm 

 

Model 

52411-17 

52411-19 

52411-20 

52411-21 

52411-18 

Electrical Requirement 

120V 60Hz 

220V 50/60Hz 

220-240V 50/60Hz 

Temperature range 

RT+10~250

 

 

Temperature stability 



 

Power consumed 

2050W 

2450W 

3100W 

4000W 

Cable Replacement 

120V 

– 12AWG 

220V 

– 1.5mm

2

/14AWG

 

1.5mm

2

/14AWG 

2.5mm

2

/12AWG 

Volume 

136L 

220L 

400L 

620L 

Fuse / Circuit breaker 

120V-20A Circuit 

Breaker, 

220V - 15A, 6 x 32 mm, 

ceramic fast blow 

16A Circuit breaker 

20A Circuit Breaker 

Interior Dimensions (W x D x H) 

550x450x550mm 

600x500x750mm 

1000x510x800 

840x600x1355mm 

 

Содержание 52411-00

Страница 1: ...52411Dryingoven_ver_1 0 Forced Air Oven Natural Convection Oven User Manual...

Страница 2: ...sed by failure to follow instructions lack of maintenance and any unauthorized modification 2 Contents of package The instrument is delivered complete with the following parts Drying Oven main unit 2...

Страница 3: ...for connecting to the mains power supply replace the plug with a suitable alternative THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN NOTE Refer to the equipment rating plate to en...

Страница 4: ...ng surface when the set temperature is over 176 F 80 C 8 Do not put hands or objects into the air inlet or air outlet 9 Unauthorized repair is not allowed The unit should have routine inspections and...

Страница 5: ...placed on solid surface to keep it horizontal 2 Aggressively opening or closing the door may cause damage to the door oven or injury to users 4 Product introductions 4 1 Function introduction Control...

Страница 6: ...52411Dryingoven_ver_1 0 www coleparmer com 4 2 Display and commands 4 2 1 Display introduction Running status Temperature Set Temperature Fan Status Buzzer MUTE ALARM Timer...

Страница 7: ...change the digit Program units tens etc of the value of the parameter you are editing 3 In combination with the key allows access to menus with password 4 and adjustment buttons allow users to increas...

Страница 8: ...aker 220V 7A 5 x 20 mm glass fast blow 120V 16A Circuit Breaker 220V 10A 5 x 20 mm glass fast blow Interior Dimensions W x D x H 250x260x250mm 340x320x320mm 420x395x350mm 450x400x450mm Model 52411 17...

Страница 9: ...uit breaker Interior Dimensions W x D x H 320x300x355mm 400x330x415mm 450x430x505mm 520x500x575mm 570x560x640mm 5 Temperature Controller Detailed Operation 5 1 Switch on 1 Connect the power according...

Страница 10: ...tion 1 After setting the operating parameters press button for about 4 5 seconds to start the heating cycle cooling process with defined time in hh mm or continue the process for 00 00 When the word E...

Страница 11: ...arts the work cycle but it doesn t immediately heat the word end and the set delay time alternately blink on the top right part of display counting the wait time until the real starting of heating Onc...

Страница 12: ...s Temperature offset on single point After set the Pb value Press to exit 6 2 Other parameters function in Menu Here are the passwords and access sequence to various parameters functions PASSWORD FUNC...

Страница 13: ...he instrument must be returned for service it is necessary to provide for proper cleaning and possible decontamination for pathogens The original package is recommended for the product if it needs to...

Страница 14: ...ing Abnormal temperature display Sensor is out or wiring is broken Repair or replace Pt100 No temperature rise Check if timing is set up and time is up Refer to the operation of timing function The co...

Страница 15: ...tific GmbH Futtererstr 16 97877 Wertheim Deutschland T 49 9377 9203 0 E cde sales antylia com W coleparmer de UK Antylia Scientific 9 Orion Court Abuscade Road Colmworth Business Park St Neots PE19 8Y...

Отзывы: