VCMC-Vacuum Calibration (continued)
Another method is to read the pressure directly from a temperature
compensated barometer that can be adjusted for altitude. This method will
give the pressure at the elevation of the calibration site without the need
for any additional calculations or corrections. Generally, the pressure value
measured should fall within the 29 to 30 inches-of-mercury range. In fact,
measurements outside this range should be suspect.
Slight variations will exist usually because of pressure variations due to
weather conditions. Therefore, the most accurate way to calibrate the
Model 282A vacuum readout system is to first obtain the atmospheric
pressure using one of the methods previously listed.
Note
The procedure below for calibrating the vacuum display assumes the
vacuum pump is being controlled by the Auxiliary Contacts, Tabs 3 & 4.
The vacuum display can be calibrated with manual pump operation, as
well.
s
To calibrate the vacuum display, perform the following:
1. While in the VCMC calibration mode, open the vacuum (full-CCW)
valve to expose the chamber to atmospheric pressure.
2. Connect a suitable vacuum pump to the PURGE port along with a
vacuum gauge as shown in the illustration below.
3. Press the SET key to enable vacuum calibration. The upper display will
show PRSR with the AM LED flashing. The lower display will indicate
29.0 in Hg.
4. Use the UP/DOWN arrow keys to set the displayed pressure to match
the atmospheric pressure.
6-14
Vacuum Oven
Cole-Parmer
Section 6
Operation
Vacuum Pump Hookup
Содержание 05017-10
Страница 1: ...Stable Temp Model 282A Vacuum Oven 105714 7006281 Rev 4...
Страница 8: ......
Страница 12: ......
Страница 36: ......
Страница 40: ......
Страница 44: ...Vacuum Oven 9 1 Cole Parmer Section 9 Electrical Schematic 115V Schematic 230V Schematic...
Страница 45: ...Cole Parmer 625 East Bunker Court Vernon Hills IL 60061 9872 1 800 323 4340 1 847 549 7600 www coleparmer com...