background image

5. Wait for the

 Ready to Calibrate

 message to display on the 

dashboard.

6. Load one optical disc in each input bin.

7. Select the component to calibrate by pressing the left or 

right arrow buttons next to 

Component

. See each section 

below for specific instructions for the Optical Drive, 
Input Bin and Printer Tray.

8. Use the movement buttons to move the robotic arm to 

various components and test their calibration settings.

Press the 

Move

 button to move the robotic arm to that 

component. If there is an optical disc on the robotic 
arm, the robotic arm will lower over the top of the 
component where it would place an optical disc in 
the component.

Press the 

Pick

 button to pick an optical disc from that 

component. If there is an optical disc on the robotic 
arm, it will not try to pick another optical disc.

Press the 

Drop

 button to place an optical disc on the 

robotic arm into that component. If there is not an 
optical disc on the robotic arm, it will not try to place 
an optical disc.

9. Adjust the calibration value until the robotic arm picks 

and places optical discs in the center of the component. 
Adjustments other than for Special Components (see 
below) can only change location right and left.

Pressing the 

right arrow

 

button will increase the value 

and move the robotic arm to the right as it would 
appear from the front of the Recorder.

Pressing the 

left arrow

 button will decrease the value 

and move the robotic arm to the left as it would 
appear from the front of the Recorder.

10. The Robotic Arm Cycle Test (see below) can be used to 

verify calibration settings.

11. If satisfied with the new calibration settings, press the 

Save

 button to accept the modified calibration values, 

exit from calibration mode and resume the normal mode 
of operation.

12. Otherwise, press the 

Cancel 

button to discard the 

modified calibration values, exit from calibration mode 
and resume the normal mode of operation.

léíáÅ~ä=aêáîÉ

The following steps define how to calibrate the picking 
position from the optical drive trays.

1. To place an optical disc onto the robotic arm, press the 

right arrow 

button or 

left arrow

 button until either 

Right 

Bin

 or 

Left Bin 

appears as the 

Component 

and then press 

Pick

2. Press the 

right arrow 

button or 

left arrow

 button until 

Bottom Drive Tray 

appears as the 

Component.

NOTE: 

Number values within the Robotic Arm Utility 

described here are different than for the external 
Recorder Calibration Utility (reference Codonics Part 
No. 901-218-001).

NOTE: 

Components (except for the Special 

Components) have calibration values between -32 and 
32 that adjust the position of the robotic arm. Left bin 
and right bin calibration values are limited to -20 
t20 for the 84-series and 101-series recorders 
only.

NOTE: 

When resuming normal operation, the software 

re-initializes the Recorder and clears any remaining 
optical discs from the drives and printer tray.

Содержание Virtua Robotic Arm Utility

Страница 1: ...the top cover and place a folded piece of paper or card stock in the optical sensor as shown in the following image This will allow access to the drives and print tray while the cover is open 3 Go to...

Страница 2: ...the Recorder Pressing the left arrow button will decrease the value and move the robotic arm to the left as it would appear from the front of the Recorder 10 The Robotic Arm Cycle Test see below can...

Страница 3: ...steps 1 through 5 7 If both optical drives need the same adjustment mark the location of the drive bay bracket on the chassis using a felt tip pen at the front of the Recorder as shown 8 Remove all op...

Страница 4: ...calibration value to move the robotic arm right and left Reselect the Pick or Move button to verify the new location 4 If the optical disc position is too far forward in the printer tray and coming c...

Страница 5: ...g the pick depth pick force 5 Use the Drop and Pick button to view how forcefully the robotic arm presses into that component adjust the force using Step 4 until satisfied 6 Repeat step 1 through 5 on...

Страница 6: ...the pick height for the print tray and DVD drives This process may be required if the pick height calibration on the GUI was attempted without a disc in the print tray or DVD drive tray 1 Connect a US...

Страница 7: ...minutes to complete 1 To begin the test press the Start button in the Robotic Arm Cycle Test section A Robot Test Running message and flashing green light will display on the dashboard A yellow light...

Отзывы: