background image

 

 

multiplying the UVB sensor reading by a factor of 40 index-cm

2

/mW, but this is 

only an estimate. The UVA sensor reading cannot be used to estimate UV index. 

Examples of experiments 

 

Measure the UV intensity as a function of time throughout the day. Do you need 
to worry about sunscreen at 8:00 in the morning?  

 

Measure the UV transmittance of various sunglasses and regular glasses. Do your 
sunglasses protect your eyes from both UVA and UVB? Can you get sunburn 
through a car window? 

 

Measure the UV transmittance of fabrics, both wet and dry. Does a wet T-shirt 
provide much solar protection? 

Calibration 

To collect data you can:  
1.

 

Use the calibration supplied by the sensor EEPROM memory. 

2.

 

Use the calibration supplied in the standard sensor library of the Coach program. 
The sensor name is 

UVA sensor (0388) (CMA) (0 .. 19700 mW/m2)

.

 

3.

 

Calibrate the sensor. It is quite difficult to calibrate a UV sensor to read in 
absolute units, since you must have a source of known UV intensity and spectral 
distribution. More often you will simply want to calibrate the sensor in terms of a 
relative intensity. In that case, you will point the sensor at a UV source (most 
often the sun) and define that intensity as 100%. To perform this kind of 
calibration, complete the following steps for a two-point calibration: 
In Coach, select the sensor you are using in the Standard Sensor Library and drag 
it to the position the sensor is connected on the panel, replacing the automatically 
detected sensor icon. Right click on the sensor icon and choose “Edit Properties” 
Set the unit to “%” and the maximum to 100.  
The first point is your zero, with no light striking the sensor. Cover the tip of the 
UV sensor with a clean opaque object. Enter 0 (zero) in “Y0” (first calibration 
point). Set “X0” of this point to the voltage that is measured (shown in the 
dialog). 
Now allow full UV intensity to strike the sensor. Since the orientation of the 
sensor affects the reading, it is best to hold the sensor in place with a ring stand 
or other clamp. To point the sensor directly at the sun, make the shadow of the 
sensor tube as small as possible. Enter 100 in “Y1” (second calibration point) and 
enter the measured voltage in “X1”. 

 

 

 

Отзывы: