Technical data 

 

  

UV peak sensitivity 

1 V per 3940 mW/m

2

 at 340 nm 

Wavelength sensitivity 

Approximate 320 to 375 nm half sensitivity points, see 
graph below for details: 

 

Resolution using 12-bit ADC 

5 mW/m

2

 

Response time 

Approximate 2 seconds to reach 95% of final reading 

Stored calibration 

I (mW/m

2

) = 3940 * V

out 

(V) 

Dimensions 

Length: 21 cm; diameter: 2 cm 

Connection 

Right-hand BT (British Telecom) connector  

 

 
 
 
 
 
Warranty:  

The 0388 UVA sensor is warranted to be free from defects in materials and 
workmanship for a period of 12 months from the date of purchase provided that it 
has been used under normal laboratory conditions. This warranty does not apply if 
the sensor has been damaged by accident or misuse.  
 

 

Note:

 

This product is to be used for educational purposes only. It is not appropriate 

for industrial, medical, research, or commercial applications.    

 

 

 

 

 

Rev. 1/11/12

 

 