4
Technical data
UV peak sensitivity
1 V per 3940 mW/m
2
at 340 nm
Wavelength sensitivity
Approximate 320 to 375 nm half sensitivity points, see
graph below for details:
Resolution using 12-bit ADC
5 mW/m
2
Response time
Approximate 2 seconds to reach 95% of final reading
Stored calibration
I (mW/m
2
) = 3940 * V
out
(V)
Dimensions
Length: 21 cm; diameter: 2 cm
Connection
Right-hand BT (British Telecom) connector
Warranty:
The 0388 UVA sensor is warranted to be free from defects in materials and
workmanship for a period of 12 months from the date of purchase provided that it
has been used under normal laboratory conditions. This warranty does not apply if
the sensor has been damaged by accident or misuse.
Note:
This product is to be used for educational purposes only. It is not appropriate
for industrial, medical, research, or commercial applications.
Rev. 1/11/12