background image

 

 

7

Do not change other options. Click OK.  

6.

 

In the same way import to the Column C4 the pressure data from the ‘Trend pressure’.  

7.

 

Open the Create/Edit Table dialog by clicking the 

Create/Edit Table

 menu option.  

 

Make data ranges C3 and C4 invisible.  

 

For C5 select ‘Formula’ and create, using the Formula Wizard, the formula [p-
pulse] – [p-trend]. Name this quantity ‘p-difference’. Make this column also 
invisible.   

8.

 

Create a new diagram with the name Envelope curve.  

 

Select for the data range C1 ‘Manual: p-trend’. 

 

 

Select for the data range C1 ‘Formula: p-difference’. 

 

An example of a bell shaped envelope curve is shown in the diagram below. 

 

 

 
 
 

 
 
 
 

 

Figure 8:

 A bell shaped “envelope curve” vs. the trend pressure. As blood pressure values 

are usually rounded to 5 mmHg we find a blood pressure of 120/80. A best-fit bell-shape 
envelope is sketched in the graph to find the maximum (MAP). 

Содержание 0377i

Страница 1: ...BLOOD PRESSURE SENSOR 0377I 0 250 MM HG User s Guide CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl Figure 1 The Blood Pressure sensor...

Страница 2: ...educational purposes Read this manual before you start measurements with the sensor Notice that over inflation of the cuff may cause pain and or injury Sensor specifications The Blood Pressure sensor...

Страница 3: ...oscillations increase with lower cuff pressure values as more blood passes through the compressed artery The maximum oscillation amplitude is reached around the mean arterial blood pressure Then as th...

Страница 4: ...ter clockwise The larger a subject s arm the slower the release valve will exhaust Helpful Tips Blood pressure readings will differ from person to person and even between measurements on the same indi...

Страница 5: ...y 50 Diastolic blood pressure is calculated using the same method and the portion for p MAP usually 75 These percentage values are obtained from empirical research 2 The bell shaped envelope curve is...

Страница 6: ...nd pressure by using the option Save as Determination of upper pressure values 1 Open the original measurement results 2 Repeat the steps 2 to 6 but this time mark the maximal pressure values the tops...

Страница 7: ...p trend Name this quantity p difference Make this column also invisible 8 Create a new diagram with the name Envelope curve Select for the data range C1 Manual p trend Select for the data range C1 For...

Страница 8: ...echnical data Pressure range 0 250 mm Hg Voltage output range 0 4 5 V Calibration function p mm Hg 64 94 Vout V Max pressure 1030 mm Hg without permanent damage Typical accuracy 3 mm Hg Temperature Co...

Отзывы: