background image

 

 

6

(displayed in the pressure 
coordinate field) is 
locally minimal. 
Select/Deselect with 
space key. In Coach 6 
you can maximize the 
Select/ Remove dialog.  

5.

 

Mark 

Keep selected

 and 

click 

OK

. As the result 

you get circa 25 points of 
the lower part of the 
envelope curve. These 
points determine the 
graph of the trend 
pressure. To get a 
smoothed graph with the 
number of points of the 
original graph through 
these points you are going to use the Spline function. Select the option 

Process > 

Smooth

6.

 

Select as Function Type 

Spline

. Type the original number of points (determined in step 

1) in the Number of Points field. Leave all other settings as they are. Mark 

Replace 

data 

and click OK. The calculations may take some time. 

7.

 

Rename the pressure quantity p to ‘p-trend’. 

8.

 

Save the results under the new name ‘Trend pressure’ by using the option 

Save as…

.  

Determination of upper pressure values  

1.

 

Open the original measurement 
results.  

2.

 

Repeat the steps 2 to 6, but this time 
mark the maximal pressure values, the 
tops of the pulses.  
It is enough to mark every second 
maximum (see the figure 7).  

3.

 

Rename the pressure quantity to ‘p-
pulse’ and save the results as ‘Pulse 
pressure’. 

Determination of the envelope curve  

You have now two result graphs from 
which the difference graph should be made. 
1.

 

Open the original measurement result. Place the measurement table in one of the Coach 
panes.  

2.

 

Select in the Table menu the 

Import data > Coach Junior/Coach 5 Result

 option. 

3.

 

Select in the dialog the result ‘Pulse Pressure’.  

4.

 

In the Import data dialog column C1 is selected in the Original data part, select in the 
Import as part ‘Column: none’. 

  

5.

 

Click column C2 in the Original data part and set in the Import as part ‘Column: C3’.  

Figure 5. 

Selection of circa 25 cuff pressure points.

Figure 7. 

Selection of the upper pressure points. 

Содержание 0377i

Страница 1: ...BLOOD PRESSURE SENSOR 0377I 0 250 MM HG User s Guide CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl Figure 1 The Blood Pressure sensor...

Страница 2: ...educational purposes Read this manual before you start measurements with the sensor Notice that over inflation of the cuff may cause pain and or injury Sensor specifications The Blood Pressure sensor...

Страница 3: ...oscillations increase with lower cuff pressure values as more blood passes through the compressed artery The maximum oscillation amplitude is reached around the mean arterial blood pressure Then as th...

Страница 4: ...ter clockwise The larger a subject s arm the slower the release valve will exhaust Helpful Tips Blood pressure readings will differ from person to person and even between measurements on the same indi...

Страница 5: ...y 50 Diastolic blood pressure is calculated using the same method and the portion for p MAP usually 75 These percentage values are obtained from empirical research 2 The bell shaped envelope curve is...

Страница 6: ...nd pressure by using the option Save as Determination of upper pressure values 1 Open the original measurement results 2 Repeat the steps 2 to 6 but this time mark the maximal pressure values the tops...

Страница 7: ...p trend Name this quantity p difference Make this column also invisible 8 Create a new diagram with the name Envelope curve Select for the data range C1 Manual p trend Select for the data range C1 For...

Страница 8: ...echnical data Pressure range 0 250 mm Hg Voltage output range 0 4 5 V Calibration function p mm Hg 64 94 Vout V Max pressure 1030 mm Hg without permanent damage Typical accuracy 3 mm Hg Temperature Co...

Отзывы: