Quickly and repeatedly squeeze the bulb to inflate the cuff on your partner’s arm. 
Continue inflating the cuff to a pressure between 150 and 170 mm Hg. The measured 
by the sensor value is displayed on the sensor icon in Coach. When the maximum 
pressure is reached, set the bulb pump down onto the table. The built-in pressure 
release valve will slowly deflate the cuff.  

5.

 

After the pressure drops to 50 mm Hg, press down on the pressure release valve to 
release any air left in the cuff. If the pressure does not reach 50 mm Hg by the time data 
collection ends, adjust the exhaust rate of the pressure release valve.  

 
Adjusting the Pressure Release Valve 

The pressure release valve is set to release at a rate of 3.0 mm Hg/second on an arm of 32 
cm in circumference. For arms much larger or much smaller it may be necessary to adjust 
the valve so that the exhaust rate stays in the range of 2.0–4.0 mm Hg/s. With the bulb in 
hand and the hose leading away from you, place a screwdriver into the metal slot on the 
top of the release valve. To increase the rate of exhaust, turn the screwdriver clockwise. To 
decrease the rate of exhaust, turn the screwdriver counter-clockwise. The larger a subject’s 
arm the slower the release valve will exhaust. 

 
Helpful Tips 

Blood pressure readings will 
differ from person to person and 
even between measurements on 
the same individual. Do not 
expect to receive the same 
measurements each trial since 
there are many factors that 
cause a person’s blood pressure 
to increase or decrease. Use the 
following tips to take accurate 
measurements.  

 

The subject’s arm and hand 
must remain still during 
measurements (hand open – 
no fist!). 

 

The arm should be at heart 
level and is best supported. 

 

Proper placement of the pressure cuff will increase the accuracy of your blood pressure 
measurements. The rubber hoses from the cuff should exit over the brachial artery and 
2 cm above the crease in the elbow. 

 

Remove any clothing that may cover or constrict the portion of the arm being 
measured. 

 

For most individuals it is not necessary to inflate the pressure cuff higher than 170 mm 
Hg. 

Over inflation of the cuff may cause pain and/or injury

.  

 

If the pressure release valve is exhausting slower or faster than 2.0–4.0 mm Hg /s, then 
adjust the exhaust rate of the pressure valve.  

 

There should be at least 10 minutes break between two measurements on the same 
individual. 

Figure 3.

 A typical measurement with the Blood Pressure 

sensor. The cuff is slowly deflated and arterial blood 
pressure pulses are detected. 

Содержание 0377i

Страница 1: ...BLOOD PRESSURE SENSOR 0377I 0 250 MM HG User s Guide CENTRE FOR MICROCOMPUTER APPLICATIONS http www cma science nl Figure 1 The Blood Pressure sensor...

Страница 2: ...educational purposes Read this manual before you start measurements with the sensor Notice that over inflation of the cuff may cause pain and or injury Sensor specifications The Blood Pressure sensor...

Страница 3: ...oscillations increase with lower cuff pressure values as more blood passes through the compressed artery The maximum oscillation amplitude is reached around the mean arterial blood pressure Then as th...

Страница 4: ...ter clockwise The larger a subject s arm the slower the release valve will exhaust Helpful Tips Blood pressure readings will differ from person to person and even between measurements on the same indi...

Страница 5: ...y 50 Diastolic blood pressure is calculated using the same method and the portion for p MAP usually 75 These percentage values are obtained from empirical research 2 The bell shaped envelope curve is...

Страница 6: ...nd pressure by using the option Save as Determination of upper pressure values 1 Open the original measurement results 2 Repeat the steps 2 to 6 but this time mark the maximal pressure values the tops...

Страница 7: ...p trend Name this quantity p difference Make this column also invisible 8 Create a new diagram with the name Envelope curve Select for the data range C1 Manual p trend Select for the data range C1 For...

Страница 8: ...echnical data Pressure range 0 250 mm Hg Voltage output range 0 4 5 V Calibration function p mm Hg 64 94 Vout V Max pressure 1030 mm Hg without permanent damage Typical accuracy 3 mm Hg Temperature Co...

Отзывы: