27/05/2020
Page 5
Operating Instructions
Usage Guidance and restrictions
•
Maximum altitude 2,000m.
•
Temperature range between 4°C and 65°C.
•
Maximum relative humidity 80% for temperatures up to 31
O
C decreas-
ing linearly to 50% relative humidity at 40
O
C.
•
Not for outdoor Use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non
-conductive pollution
occurs.
Occasionally, however, a temporary conductivity caused by condensation
must be expected”.
Setting up the Cellulose Acetate Tanks
Fitting Electrode cables
•
Note the position of the lid on the unit. This shows the correct polarity
and the correct orientation of the cables, black is negative and red
positive.
•
Remove the lid from the unit. Note if the lid is not removed, fitting the
cables may result in un-tightening of the gold plug and damage to the
electrode.
•
Screw the cables into the tapped holes as fully as possible so that there
is no gap between the lid and the leading edge of the cable fitting.
•
Refit the lid.
The unit is now ready to be used.