Operating Instructions 

Usage Guidance and restrictions 

 

Maximum altitude 2,000m. 

 

Temperature range between 4°C and 65°C. 

 

Maximum relative humidity 80% for temperatures up to 31

O

C decreas-

ing linearly to 50% relative humidity at 40

O

C. 

 

Not for outdoor Use. 

This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664. 

POLLUTION DEGREE 2, states that: “Normally only non

-conductive pollution 

occurs. 
Occasionally, however, a temporary conductivity caused by condensation 

must be expected”.

 

Setting up the Cellulose Acetate Tanks 

Fitting Electrode cables 

 

Note the position of the lid on the unit. This shows the correct polarity 
and the correct orientation of the cables, black is negative and red 
positive.  

 

Remove the lid from the unit. Note if the lid is not removed, fitting the 
cables may result in un-tightening of the gold plug and damage to the 
electrode. 

 

Screw the cables into the tapped holes as fully as possible so that there 
is no gap between the lid and the leading edge of the cable fitting. 

 

Refit the lid. 

The unit is now ready to be used.

 

 

Содержание CSL-CELLAS

Страница 1: ...CELLASMINI Record the following for your records Model _____________________ Catalogue No _____________________ Date of Delivery _____________________ Warranty Period _____________________ Serial No _...

Страница 2: ...ng List 4 Operating Instructions 5 Usage Guidance and restrictions 5 Setting up the Cellulose Acetate Tanks 5 Fitting Electrode cables 5 Preparation and Gel running 6 Example of Staining 8 Example of...

Страница 3: ...e complete manual thoroughly The unit must never be used without the safety lid correctly in po sition The unit should not be used if there is any sign of damage to the exter nal tank or lid These uni...

Страница 4: ...units are received to ensure that all components have been included The unit should be checked for damage when received Cleaver Scientific is liable for all missing or damaged parts accessories within...

Страница 5: ...owever a temporary conductivity caused by condensation must be expected Setting up the Cellulose Acetate Tanks Fitting Electrode cables Note the position of the lid on the unit This shows the correct...

Страница 6: ...sition the electrode wicks in the tank so that they overhang the wick carrying bars with the base sit ting in the buffer tanks see diagram below Filter paper or paper towels can be used as the wicks 3...

Страница 7: ...ply the samples at the negative end of the cellulose acetate plate Usually a small amount of dye can also be loaded to indicate sample migration distance The samples should be kept as cold as possible...

Страница 8: ...hly rinsed with warm water or distilled water to prevent build up of salts but care should be taken not to damage the en closed electrode and vigorous cleaning is not necessary or advised Air drying i...

Страница 9: ...ent as described above Wash with 3 hydrogen peroxide H2O2 for 10 minutes Rinsed with 0 1 DEPC diethyl pyro carbonate treated distilled water Caution DEPC is a suspected carcinogen Always take the nece...

Страница 10: ...ne other than CSL or an appointed distributor or representative are no longer under warranty from the time the unit was modified Units which have accessories or repaired parts not supplied by CSL or i...

Отзывы: