Retighten the four 5mm screws.
Check the pin-jaw alignment for pin concentricity.
With the air turned off and the
magazine in, place a piece of tubing in the jaw set and clamp the tubing manually with
the jaws. Pull the pin forward manually through the magazine and ensure that it lines up
with the center of the inner diameter of the tubing. Gently rotate the pin to ensure that
there is no wobble in the pin head.
Failure to do so may result in severe damage to both the magazine and jaw set!
If any wobble is detected, recheck that the threads on the cylinder and pin set are clean.
If there is still a wobble, the Pin Set Assembly is bad and another must be used.
2.5 Adjusting the Stop Gate
The distance between the end of the jaw set and the stop gate may need to be adjusted
according to what fitting is being used. This distance to adjust the stop gate is
determined by the gap between the fitting flange and the end of the tubing. Sliding the
stop gate away from the front of the unit will insert the fitting deeper into the tubing.
Figures 6 and 7 below show the tubing before adjusting the stop gate and after
adjusting the stop gate:
Figure 6 Figure 7
Unadjusted Stop Gate Adjusted Stop Gate
To adjust this distance, loosen the one 5mm socket cap screw on the side of the gate
and move the gate in or out depending on the stick out desired (figures 8 and 9 below).
Re-tighten the screw before attempting to run the unit.