Cr 2021-02-01 vs. 2.3.
User Manual for VCS-20Cr
24
Steps: Next, Next, Install, Finish
Create network connection to VCS-20Cr device with cable or WLAN.
Ending operation
When the targeted time is reached, the process shuts down.
Do not switch off the device while the cooling indication is still in the screen.
▪
The hydrogen peroxide concentration in the decontaminated space must be
less than 1 PPM, before entry without the necessary protective equipment.
The concentration must always be checked using a detector.
▪
The decomposition of hydrogen peroxide vapour can be accelerated with an
external catalytic converter.
▪
The ventilation of the space must be planned case-specifically and in
accordance with valid regulations. Note that porous materials within the
decontaminated space may increase the hydrogen peroxide aeration time.
▪
Switch off the device, detach the power cord by lifting the latch and turning
the socket counter-clockwise until it gets loose
▪
Detach the hydrogen peroxide sensor by lifting it up with some force.
▪
Remove the hydrogen peroxide bottle.
▪
Close the case.
Detaching the hydrogen peroxide bottle
1. Detach both of the suction hoses first from the unit and next from the bottle.
2. Remove the ventilated cap from the bottle and replace it with a cap for storage.
3.
If the bottle is empty rinse it with clean, cold water.
4.
Use appropriate personal protective equipment (PPE).