Cr 2021-02-01 vs. 2.3.
User Manual for VCS-20Cr
15
Self check
I In
Self check
the device will test all the
critical parts of the device. It is possible to
skip this phase with pressing
Settings
button.
However, the Sensor will do its
own 6 minutes start-up calibration
procedure.
S Set the
Amount of H2O2
in millilitres. Set the cubic meters of the
Target Size
. Set the
desired PPM level on
Set Point
. With
Power-%
parameter you can adjust the vapor
generating.
The amount of H2O2 can be
RESET.
The amount left will move to the field
Amount of
H2O2 (ml).
The
amount of H2O2-
value
cannot be 0.
Supply
All the buttons with a picture
of hourglass needs to be
pushed for 2 seconds.
Edit the treatment values on
Set
page.
Set the desired
Treatment time
. Set the
Target PPM level.
Specify
RS value to field
RS Limit
%.
Set the lowest PPM level where the treatment time will start using the field
Start treatment
time at
.
The
Wi-Fi
is
ON
, when the background of the button is dark. When Wi-Fi is disabled,
the WiFi-card is un-powered so that the device cannot be hacked. This might be needed in
high security areas.
Set