SPECIFICATIONS

FIELD OF VIEW

Temperature range: -50°C to 1350°C (-58°F to 2462°F)

Accuracy :

Range

Accuracy

Celsius

-50°C to -20°C

+/- 5°C

-20°C to 200°C

+/- (1.5% of r 2°C)

200°C to 538°C

+/- (2% of r 2°C)

538°C to 1350°C

+/- (3% of r 5°C)

Fahrenheit

-58°F to -4°F

+/- 9°F

-4°F to 392°F

+/- (1.5% of r 3.6°F)

392°F to 1000°F

+/- (2% of r 3.6°F)

1000°F to 2462°F

+/- (3% of r 9°F)

As the distance (D) from the abject being measured increases, the spot size (S) of the area measured by the 

thermometer becomes larger. This relationship between distance and spot size is normally expressed as the distance 

to spot, or D: S ratio, which is shown in the following fi gure. At a distance of 12 inches, the spot size would be 1 

inch in diameter. The thermometer will display the average temperature across the spot area.

Note :

Make sure that the object tobe 

measured is larger than the 

thermometer's spot size. The smaller 

the object, the closer you should be 

to it. 

When accuracy is critical, make sure 

that the object is at least twice as 

large as the spot size. 

Avoid exposure

Laser radiation is emitted through the aperture

Output <1mW

Longueur d’onde 630nm – 670nm

Cautions

The thermometer should be protected from the following:

1. Electre magnetic fi elds ( created by arc welders, induction heaters and similar items. )

2. Thermal shock ( caused by large or abrupt ambient temperature changes. Allow 30 minutes for the thermometer 

to stabilize before use.)

3. Do not leave the thermometer on or near abjects of high temperature.

Содержание OE 5027

Страница 1: ...clas equipements com THERMOMETRE INFRAROUGE 50 1350 INFRARED THERMOMETER 50 1350 OE 5027...

Страница 2: ...ctionne le thermom tre Le thermom tre infrarouge est un syst me optique d tecteur photo amplificateur de signal traitement du signal et syst me de visualisation Le syst me optique collecte l nergie du...

Страница 3: ...e A une distance de 30cm la dimension du point sera de 2 50cm de diam tre Le thermom tre affichera la temp rature moyenne de la zone du point Note S assurer que l objet mesurer est plus large que le p...

Страница 4: ...nt d utilisation Temp rature 0 C 40 C Taux d humidit 10 95 RH sans condensation Temp rature de stockage 20 C 50 C Poids 325g batterie incluse Dimension 20 7 x 19 5 x 5 2 cm Batterie 9V batterie 6F22 o...

Страница 5: ...ur affiche maximum MAX minimum MIN diff rence DIF moyenne AVG alarme de valeur haute HAL et alarme de valeur faible LAL 9 HI est l ic ne pour l alarme forte et LOW est l ic ne de l alarme faible 10 In...

Страница 6: ...m moire et le thermom tre peut commencer un nouvel enregistrement Quand vous appuyez sur la g chette la lecture sur l cran principal va changer la temp rature de la surface cibl e quand vous rel cher...

Страница 7: ...s que le thermom tre est allum appuyer sur le bouton MODE jusqu ce que le symble E s affiche puis rel cher le bouton 2 Appuyer sur le bouton MODE jusqu ce que LAL apparaisse maintenant le niveau d ala...

Страница 8: ...ige 0 83 plastique 0 85 0 95 vitre 0 90 0 95 bois 0 90 c ramique 0 90 0 94 papier 0 70 0 94 marbre 0 94 Oxyde de chrome 0 81 pl tre 0 80 0 90 Oxyde de cuivre 0 78 mortier 0 89 0 91 Oxyde de fer 0 78 0...

Страница 9: ...l eau ne pas laisser entrer de liquide dans le carton Remplacement de la batterie Lorsque le symbole appara t la batterie est faible et doit tre remplac e imm diatement Attendre que le thermom tre s...

Страница 10: ...eter works The infrared thermometer consists of optical system photo detector signal amplifier signal processing and display output system and etc The optical system collects the infrared radiation en...

Страница 11: ...following figure At a distance of 12 inches the spot size would be 1 inch in diameter The thermometer will display the average temperature across the spot area Note Make sure that the object tobe mea...

Страница 12: ...0 C to 40 C Relative humidity 10 to 95 RH noncondensing up to 30 C Storage Temperature 20 C to 50 C Weight about 325g including battery Size 20 7 X 19 5 X 5 2cm Battery 9V battery 6F22 or equivalent 1...

Страница 13: ...display is maximum value 8 Bottom display used to display maximum MAX minimum MIN difference DIF average AVG high alarm value HAL and low alarm value LAL 9 Hl is high alarm icon and LOW is low alarm i...

Страница 14: ...erased from memory and the thermometer will start new recording When you pull the trigger the reading on the main display will change with the temperaturer of the target surface and when yourelease t...

Страница 15: ...through transparent surfaces such as glass lt will measure the surface temperature of the glass instead 3 Steam dust smoke can prevent accurate measurement by interfering with the energy emitted from...

Страница 16: ...in 0 98 Cement 0 96 Lather 0 75 0 80 Sand 0 90 Charcoal powder 0 96 Earth 0 92 0 96 Lacquer 0 80 0 95 Water 0 92 0 96 Laquer matt 0 97 Ice 0 96 0 98 Rubber black 0 94 Snow 0 83 Plastic 0 85 0 95 Glass...

Страница 17: ...en the symbol appears the battery is low and should be replaced immediately Wait untilthe thermometer turns off automatically then open the battery cover gently with thumb at the location indicated in...

Страница 18: ...NOTES...

Страница 19: ...NOTES...

Страница 20: ...le technicien de votre distributeur agr If you need components or parts please contact the reseller In case of problems please contact your authorized technician CLAS Equipements ZA de la CROUZA 73800...

Отзывы: