OE 3004
DIGITAL MULTIMETER CAT. III
DC CURRENT
Range
Resolution
Accuracy
Overage indication
20mA
10µA
± (1.0% + 5)
Only fi gure "1" shown on
the display
200mA
100µA
± (1.5% + 5)
10A
10mA
± (2.0% + 5)
-- [1]
Overload Protection:
250mA/300V FAST Fuse (for "ΩVmA°C" terminal inputs) 10A/300V FAST Fuse (for "10A" terminal inputs)
Max. Allowed Input Current: 1OA
( For inputs > 2A: measurement duration < 10 secs, and interval > 15 minutes )
[1] If the current being measured is > 10A, the display may
show the value of the current; but the measurement is dangerous.
AC CURRENT
Range
Resolution
Accuracy
Overage indication
20mA
10µA
± (1.3% + 5)
Only fi gure "1" shown on
the display
200mA
100µA
± (1.8% + 5)
10A
10mA
± (3.0% + 5)
-- [1]
Overload Protection:
250mA/300V FAST Fuse (for "ΩVmA°C" terminal inputs) 10A/300V FAST Fuse (for "10A" terminal inputs)
Max. Allowed Input Current: 1OA
(For inputs > 2A: measurement duration < 10 secs, and interval > 15 minutes)
Frequency Range: 40Hz - 400Hz
Response: Average, calibrated in rms of sine wave
[1] If the current being measured is > 10A, the display may show the value of the current; but the measurement is
dangerous.
RESISTANCE
Range
Resolution
Accuracy
Overage indication
200Ω
0.1Ω
± (1.2% + 5
Only fi gure "1" shown on
the display
2kΩ
1Ω
20kΩ
10Ω
200kΩ
100Ω
2MΩ
1kΩ
20MΩ
10kΩ
± (1.5% + 7)
Max. Open Circuit Voltage: About 2.8V