- Eng 25 -
English
ELECTOROMAGNETIC COMPATIBILITY INFORMATION
WARNING
• Portable RF communications equipment should be used no closer than 30 cm (12
inches) to any part of the [CH-432], including cables specified.
• Use of this equipment adjacent to or stacked with other equipment should be avoided.
• Use of accessories and options other than those specified (other than CITIZEN
original parts) could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment.
Guidance and manufacturer’s declaration - electromagnetic emissions
The [CH-432] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [CH-432] should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The [CH-432] uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The [CH-432] is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
N/A
Guidance and manufacturer’s declaration - electromagnetic immunity
The [CH-432] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [CH-432] should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output lines
N/A
N/A
Surge
IEC 61000-4-5
±1 kV
line to line
±2 kV
line to earth
N/A
N/A
Voltage dips,
short
interruptions
and voltage
variations on
power supply
IEC 61000-4-11
0%
U
T
0.5 cycle
N/A
N/A
0%
U
T
1 cycle
70%
U
T
25/30 cycle
0%
U
T
250/300 cycle
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note:
U
T
is the A.C. mains voltage prior to application of the test level.
02̲CH-432̲1901̲Eng.indd 25
2019/06/03 13:31
Содержание CH-432
Страница 8: ...Hin 8 01_CH432_Hin indd 8 01_CH432_Hin indd 8 3 6 2019 1 56 25 PM 3 6 2019 1 56 25 PM...
Страница 9: ...Hin 9 01_CH432_Hin indd 9 01_CH432_Hin indd 9 3 6 2019 1 56 25 PM 3 6 2019 1 56 25 PM...
Страница 10: ...Hin 10 1 2 4 AA 01_CH432_Hin indd 10 01_CH432_Hin indd 10 3 6 2019 1 56 25 PM 3 6 2019 1 56 25 PM...
Страница 11: ...Hin 11 LR6 R6P 3 01_CH432_Hin indd 11 01_CH432_Hin indd 11 3 6 2019 1 56 25 PM 3 6 2019 1 56 25 PM...
Страница 13: ...Hin 13 4 5 6 01_CH432_Hin indd 13 01_CH432_Hin indd 13 3 6 2019 1 56 25 PM 3 6 2019 1 56 25 PM...
Страница 14: ...Hin 14 280mmHg 01_CH432_Hin indd 14 01_CH432_Hin indd 14 3 6 2019 1 56 25 PM 3 6 2019 1 56 25 PM...
Страница 15: ...Hin 15 01_CH432_Hin indd 15 01_CH432_Hin indd 15 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...
Страница 17: ...Hin 17 10mmHg 01_CH432_Hin indd 17 01_CH432_Hin indd 17 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...
Страница 18: ...Hin 18 3A 3A 01_CH432_Hin indd 18 01_CH432_Hin indd 18 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...
Страница 20: ...Hin 20 40 181 PUL Err 281mmHg 01_CH432_Hin indd 20 01_CH432_Hin indd 20 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...
Страница 22: ...Hin 22 10 20 mmHg 01_CH432_Hin indd 22 01_CH432_Hin indd 22 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...
Страница 23: ...Hin 23 22 01_CH432_Hin indd 23 01_CH432_Hin indd 23 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...
Страница 25: ...MEMO 01_CH432_Hin indd 25 01_CH432_Hin indd 25 3 6 2019 1 56 26 PM 3 6 2019 1 56 26 PM...