6
Model 008A, 18023-A & 18043-A Assembly Procedure
1.
Place base plate with actuator assembly
on table. Carefully move base plate to
overhang table. This allows access to
holes for attaching phantom body.
2.
Position phantom body on base plate as
shown. The rod hole must be on same
side and aligned with the rod holder.
3.
Place one screw through bottom of base
plate and screw into phantom. Continue
until all screws are in place. Snug but do
not tighten screws to allow minor align-
ment adjustments in steps 6 and 7.
4.
Insert lung rod thru hole in phantom
body. Ensure that screw hole on lung rod
is aligned with screw hole on actuator
rod.
5.
Use thumb screw to secure lung rod to
actuator rod.
6.
Push lung rod towards the motor box.
Visually check gap around the rod. It
should appear consistent. If not, adjust
body left and right until gap is consistent
around rod.
7.
Repeat step 6 with rod extended out.
Check gap on the motor box side of
body as well. Adjust phantom body to
achieve best fit. Finish tightening the
screws on underside of phantom.
8.
3rd axis actuator is shipped in the
horizontal position.
9.
Move 3rd axis gating device to vertical
position by loosening screws and
repositioning as shown.
CAUTION:
PHANTOM ASSEMBLY REQUIRES 2 PEOPLE. ONE PERSON FOR STABILIZING AND A SECOND FOR ASSEMBLING THE PHANTOM.
FAILURE TO STABILIZE THE BASE PLATE CAN RESULT IN DAMAGE TO THE PHANTOM.