Cirris Touch1 LV Скачать руководство пользователя страница 5

Touch1 LV Checkout Manual -  Page 3 

 

 

Set Up 

 

 

 

 

Packing List: 

 

You should have received the following items with your 

Touch1 LV Performance Verification 

Kit

  This manual which includes a copy of a Touch1 LV Certificate of Calibration 

  0.1% Resistor Leak Adapter 

  T1/H+ Capacitance/4-Wire Adapter 

  Zero Ohm Adapter 

 

 

 

 

 

Tester Preparation: 
 

To prepare the tester for calibration, do the following: 

 

• 

  Install any expansion boxes you want to use. For information on how to do this, see 

the online help system or your users manual. 

Note: The examples in this manual show a Touch1 LV with no expansion boxes. 

• 

  Connect the power cord to the tester. Plug the tester into a grounded power outlet. 

• 

  To print Tester Checkout results, connect a printer to the tester using the parallel port. 

• 

 

Depending on your calibration system requirements, you may need to prepare a form 
to record the results of the calibration. At the end of this manual is a 

Touch1 LV 

Certificate of Calibration

. If you use this form, make a photocopy so you can maintain 

the master for the next time you perform the calibration tests.

 

Содержание Touch1 LV

Страница 1: ...Touch1 LV Performance Verification Manual Version 1 0 January 28 2010 ISO 9001 Certified...

Страница 2: ...erformance Verification Manual Version 1 0 Editor Karen Johnson Copyright 2004 by Cirris Systems Corporation All Rights Reserved Cirris Systems Corporation 1991 W Parkway Boulevard Salt Lake City Utah...

Страница 3: ...ble of Contents Introduction 2 Set Up 3 Tester Checkout Tests 4 Self Test Touch Pad Test Probe Test Low Voltage Threshold Tests Capacitance and 4 Wire Tests Scanner Tests Tester Checkout Summary 4 App...

Страница 4: ...th instruments traceable to the National Institute of Standards and Technology NIST Calibration should be performed on your Touch1 LV tester at least once a year You should also check the calibration...

Страница 5: ...tion on how to do this see the online help system or your users manual Note The examples in this manual show a Touch1 LV with no expansion boxes Connect the power cord to the tester Plug the tester in...

Страница 6: ...If there is no checkmark inside the button the test is turned off To start a test press Next Test X where X lists the test to be performed The Touch1 will perform the test and then show either a Pass...

Страница 7: ...the voltage measurement gain circuitry Test 4 Calibrates most of the current sources Test 5 Not applicable for R type scanners Test 6 Checks the scanners Test 7 Checks the scanner test points Test 8 V...

Страница 8: ...er of the picture inside the button 4 A screen with five buttons will be displayed Press each button to check the calibration You must press a button within ten seconds or the test will fail A star sh...

Страница 9: ...e probe to any test point s on the adapter If the probe is working correctly a pop up window will display the correct point s touched by the probe If no pop up window is displayed when the test point...

Страница 10: ...he Performance Verification Kit is needed to perform this test 1 To start the test press Next Test LV HV Meas 2 A screen will prompt you to install the Resistance Adapter in the J1 position on the sca...

Страница 11: ...ce 4 Wire Adapter included in the Performance Verification Kit is needed to perform this test 1 To start the test press Next Test Cap 4W 2 A screen will prompt you to install the Capacitance 4 Wire Ad...

Страница 12: ...of the problem The Zero Ohm Adapter included in the Performance Verification Kit is needed to perform this test 1 To start the test press Next Test Scanners 2 A screen will prompt you to install the...

Страница 13: ...apter was removed you must press the NO RESPONSE button Remove the adapter to automatically proceed to the next test 5 Remove the Zero Ohm Adapter and place it in the J3 J4 position as shown 6 If the...

Страница 14: ...e tests a scanner summary screen will be displayed listing the results for each test section If any of the the individual scanner tests do not pass the entire ScannerTest will fail Results for two sep...

Страница 15: ...display SKIPPED for any skipped tests The summary will display FAILED for any failed tests 2 To save the Test Summary results you must either print them or copy them to paper by hand If you have a pr...

Страница 16: ...tional standards for measurements in the United States Good Calibration Practices The calibration standards such as ANSI NCSL Z540 1 and ISO 10012 1 require several good practices for the calibration...

Страница 17: ...can be used as a calibration data report This report can be printed or copied to paper by hand Traceability Did qualified personnel perform the calibration procedure under controlled conditions using...

Страница 18: ...umber Calibrated by Calibration Date Due Date Applicable Calibration Standard s Procedure Touch1 LV Performance Verification Version 1 0 Temperature Relative Humidity Tester Serial Number Instruments...