background image

Touch1 LV Checkout Manual -  Page 14 

 

 

Appendix 

 

 

 

 
The following information can be used a guide for setting up a formal calibration system in 
your organization. 

Calibration Standards 

Calibration standards refer to written quality system requirements for organizations that 
perform calibations and use calibrated equipment. Establishing a quality system according 
to calibration standards helps insure calibrations are done competently and lends credibility 
to the calibration organization. In the United States, common calibration standards include 
ANSI/NCSL Z540-1, ISO/IEC Guide 25, ISO 10012-1, and the former MIL-STD 45662A.  

 

The ANSI/NCSL Z540 standard refered to above, as well as other helpful metrology 
information, can be obtained from the National Conference of Standards Laboratories 
International (NCSL) at 1-303-440-3339 or 

www.ncslinternational.org

. You can obtain the 

ISO standards from the International Standards Organization (ISO) at their web site 

www.iso.net/

 

The word “standards” is often used in another way in the metrology industry to refer to a 
centralized, most accurate unit of measurement regulated by countries.  The National 
Institute of Standards and Technology (NIST) maintains the 

national standards

 for 

measurements in the United States.

 

Good Calibration Practices 

The calibration standards such as ANSI/NCSL Z540-1 and ISO 10012-1 require several 
good practices for the calibration industry including the following areas:

 

 Recall 

System 

How do you ensure you do not forget to send an instrument in for calibration? Use a  
recall system which can be a card file or a computerized database. This system 
includes calibration dates, due dates, calibration sources, and other instrument 
records. The recall system ensures calibrated instruments are recalibrated in a timely 
manner.

 

 Calibration 

Labels

 

How does someone know if an instrument has been calibrated without looking for the 
paperwork in a filing cabinet drawer? When an instrument is calibrated, the calibration 
standards require the instrument to be labeled as such. The calibration labels, which 
are applied to instruments, have fields for the instrument serial number, calibration 
date, calibration due date, and by whom. A good source of inexpensive calibration 
labels is United Ad Label at 1-800-992-5755.

 

 Accuracy 

Ratios

 

Can you use a ruler to calibrate your digital calipers? Of course not! Wherever 
possible calibration standards require an accuracy ratio of at least four to one. In 
other words, the insturment being used to measure the calibrated instrument must be 
at least four times as accurate as the calibrated instrument.

 

Содержание Touch1 LV

Страница 1: ...Touch1 LV Performance Verification Manual Version 1 0 January 28 2010 ISO 9001 Certified...

Страница 2: ...erformance Verification Manual Version 1 0 Editor Karen Johnson Copyright 2004 by Cirris Systems Corporation All Rights Reserved Cirris Systems Corporation 1991 W Parkway Boulevard Salt Lake City Utah...

Страница 3: ...ble of Contents Introduction 2 Set Up 3 Tester Checkout Tests 4 Self Test Touch Pad Test Probe Test Low Voltage Threshold Tests Capacitance and 4 Wire Tests Scanner Tests Tester Checkout Summary 4 App...

Страница 4: ...th instruments traceable to the National Institute of Standards and Technology NIST Calibration should be performed on your Touch1 LV tester at least once a year You should also check the calibration...

Страница 5: ...tion on how to do this see the online help system or your users manual Note The examples in this manual show a Touch1 LV with no expansion boxes Connect the power cord to the tester Plug the tester in...

Страница 6: ...If there is no checkmark inside the button the test is turned off To start a test press Next Test X where X lists the test to be performed The Touch1 will perform the test and then show either a Pass...

Страница 7: ...the voltage measurement gain circuitry Test 4 Calibrates most of the current sources Test 5 Not applicable for R type scanners Test 6 Checks the scanners Test 7 Checks the scanner test points Test 8 V...

Страница 8: ...er of the picture inside the button 4 A screen with five buttons will be displayed Press each button to check the calibration You must press a button within ten seconds or the test will fail A star sh...

Страница 9: ...e probe to any test point s on the adapter If the probe is working correctly a pop up window will display the correct point s touched by the probe If no pop up window is displayed when the test point...

Страница 10: ...he Performance Verification Kit is needed to perform this test 1 To start the test press Next Test LV HV Meas 2 A screen will prompt you to install the Resistance Adapter in the J1 position on the sca...

Страница 11: ...ce 4 Wire Adapter included in the Performance Verification Kit is needed to perform this test 1 To start the test press Next Test Cap 4W 2 A screen will prompt you to install the Capacitance 4 Wire Ad...

Страница 12: ...of the problem The Zero Ohm Adapter included in the Performance Verification Kit is needed to perform this test 1 To start the test press Next Test Scanners 2 A screen will prompt you to install the...

Страница 13: ...apter was removed you must press the NO RESPONSE button Remove the adapter to automatically proceed to the next test 5 Remove the Zero Ohm Adapter and place it in the J3 J4 position as shown 6 If the...

Страница 14: ...e tests a scanner summary screen will be displayed listing the results for each test section If any of the the individual scanner tests do not pass the entire ScannerTest will fail Results for two sep...

Страница 15: ...display SKIPPED for any skipped tests The summary will display FAILED for any failed tests 2 To save the Test Summary results you must either print them or copy them to paper by hand If you have a pr...

Страница 16: ...tional standards for measurements in the United States Good Calibration Practices The calibration standards such as ANSI NCSL Z540 1 and ISO 10012 1 require several good practices for the calibration...

Страница 17: ...can be used as a calibration data report This report can be printed or copied to paper by hand Traceability Did qualified personnel perform the calibration procedure under controlled conditions using...

Страница 18: ...umber Calibrated by Calibration Date Due Date Applicable Calibration Standard s Procedure Touch1 LV Performance Verification Version 1 0 Temperature Relative Humidity Tester Serial Number Instruments...

Отзывы: