6.2.4. “Pass / Fail” Test mode
Press
key in order to determine the lower insulation limit for type
“Pass / Fail” test. Select this value using
and
keys. Possible
values are 10 MΩ, 100 MΩ, 1 GΩ
or 10 GΩ.
During a “Pass/ Fail” test, the
MD-5060e
will indicate when the insulation
resistance is lower than the programmed limit, both with an intermittent
beep and the
key led flashing. The
key led will remain flashing
until the end of tests, or until the measurement of the resistance value is
greater than the programmed limit.