----- Analyzer of power supply quality CAVA-251 / CAVA-252 / CAVA-253 -------  Page  No.    12 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

7.- OPERATION MODE 

 

Before connecting the analyzer to a single-phase power system in order to 

record data or to transfer data to a P.C. check the “Hold/Run” switch to be at the right 
position. 

The position of this switch is only considered when powering up the 

analyzer.

 No later modification of the switch position will be effective once the 

analyzer is already running. 
 
 

It is advisable to observe the “Hold / Run” led to check that the analyzer is at 

the right operation mode according to our requirements (See section 6.2.-)  
 

The position of this switch is only considered when powering 

up the analyzer. 

 

7.1.- Communications 
 

Setup procedure and data retrieval can only be performed through the RS-232 

serial port in the CAVA; therefore, the configuration of the serial port of the PC and 
the CAVA must entirely coincide. 

 
The default configuration of the CAVA serial port is: 

 
Peripheral  : 00    

 

Communication parameters:  9.600 / 7 / N / 1 

 

The CAVA only permits to modify the baud rate. Available baud rates are: 2400, 

4800, 9600 & 19200 bauds 
 

When setting the data management software for the CAVA in the PC, the serial 

port used to communicate to the CAVA (Com 1, Com 2) must be correctly defined. 
The software will automatically check at different baud rates whether a CAVA is 
connected to the set port. 

Содержание CAVA-251

Страница 1: ...ANALYZER OF POWER SUPPLY QUALITY CAVA Series 251 252 253 USER S MANUAL M98120201 03 03A c CIRCUTOR S A...

Страница 2: ...measured and calculated 6 4 2 Parameters referred Power Supply Quality 7 4 2 1 Harmonics 7 4 2 2 Flicker 7 4 2 3 Voltage sags 8 4 2 4 Voltage dropouts Error Marcador no definido 4 2 5 Percentage of V...

Страница 3: ...ntrol of the quality of L V power systems You are kindly requested to carefully read this manual before connecting and powering the analyzer in order to avoid irreversible damage which might be caused...

Страница 4: ...tion instructions Before powering and connecting the analyzer check the following points a Supply voltage 230 V a c 15 b Frequency 50 60 Hz c Maximum measurable current according to the ammeter clamp...

Страница 5: ...d by means of a P C Before powering the analyzer please carefully read sections referred to CONNECTIONS AND SET UP to choose the most appropriate operation mode for your requirements The CAVA measures...

Страница 6: ...ltage sags and voltage dropouts Measurement of harmonic distortion THD or D Measurement of Flicker 3 CAVA MODELS Different CAVA models are available according to measured and recorded parameters CAVA...

Страница 7: ...Vn Vrms 1 2 T u dt t Vrms 1 2 1 N u N CURRENT RMS value CAVA 252 CAVA 253 In Irms 1 2 T i dt t Irms 1 2 1 N i N ACTIVE POWER CAVA 253 Pn 1 T u t i t dt P 1 1 N u i N POWER FACTOR of each phase PF1 PF...

Страница 8: ...ty Pst RMS Flicker WA Value of the Flicker RMS that is defined as the average result of the weighted Flicker value due to each frequency N n n N weighted Flic FT V dV n Flic RMS 1 2 2 1 ker ker Severi...

Страница 9: ...s The CAVA is also equipped with a counter to give the number of cycles within the last period where a voltage dropout has occurred The CAVA will also provide the number of intervals these voltage dro...

Страница 10: ...x x Current x x Apparent power x Active power x Reactive power x Power Factor x Electric parameters Frequency x x x No of cycles x x x Voltage sags Voltage dropouts Intervals x x x No of cycles x x x...

Страница 11: ...ected to exist or external damages are observed the instrument must be immediately put out of service Contact then with a qualified service representative 6 1 Installing the CAVA Before powering and c...

Страница 12: ...mp otherwise no power value will be recorded by the CAVA 253 Hold Run led This led indicates several operation modes Whether the led is quickly blinking a reset action is being carried out Whether the...

Страница 13: ...s See section 6 2 The position of this switch is only considered when powering up the analyzer 7 1 Communications Setup procedure and data retrieval can only be performed through the RS 232 serial por...

Страница 14: ...fault to provide the average value of the voltage for a period of one second If required this integration time can be modified just setting at this option the desired period is seconds Voltage thresho...

Страница 15: ...d frequency The rated frequency of the monitored power system must be set to obtain right data when measuring instantaneous values Time constant for frequency measurement CAVA 252 CAVA 253 The CAVA is...

Страница 16: ...ture range 0 to 50 C Current measurement according to ammeter clamp Accuracy class Voltage 0 5 or readout Current 0 5 or readout Powers 1 or readout Test conditions for accuracy determination Ammeter...

Страница 17: ...nce or repairing action should be done over the instrument open and powered and should those actions are essential high qualified operators must perform them Before any adjustment replacement maintena...

Отзывы: