2.5.2
Sending a Ramp/Soak Program to the EZT-570S
Programs are sent to the EZT in a step-by-step process. The download sequence must be followed
in order and must complete without errors to be valid. If a write error is detected during the transfer of
a program from a PC to the EZT (no response from EZT or NACK returned), the program download
must be aborted and restarted.
The EZT-570S is put into program transfer mode when the first group of registers containing the
program specific data is sent (registers 200-214). The EZT then begins looking for the number of
steps of the program to be sent as was set in register 209. As each step is received, it increments
the count. Once all steps have been received, the EZT downloads the program into the control
module memory. During this transfer, register 180 will be set to 1 to indicate that the process is
taking place. Once the register value returns to zero, the program is ready to be started.
If an error occurs during the transfer process from the PC to the EZT, the program transfer process
should be stopped at the PC. The data sent to the EZT was either corrupted in transmission or not
received properly. It is not possible to resend the failed step because it is not known if any of the
previous data was received by the EZT properly. On the transmission error, the EZT will enter a 15
second timeout process. At the end of the timeout period, the buffer will be cleared and the program
can be resent. In order to insure that the new download begins properly, induce a 20 second wait
period on the host PC after the failed transmission attempt to insure that enough time has elapsed.