EZT-570S User Communication Reference Manual
18
2.4 EZT-570S
Control
Registers
The EZT-570S is capable of utilizing up to five control loops and eight monitor inputs. The register list
in this section of the manual lists the associated values for all of the loops, inputs and their associated
alarms by the loop or monitor input number, i.e., 1 - 5 and 1 - 8. While the monitor inputs will be easy
to decipher, since they are shipped from the factory with the relative number in their tag name, the
loops are not. The loop names are defined by the chamber process they control, i.e., temperature,
humidity, etc., thus the number of control loops required and their function can vary between different
chamber models.
The EZT-570S displays all control loops and monitor inputs in sequential order. The loop/monitor
order can be viewed from the “All Loops View” screen. Starting at the top of the list and counting
down, the first entry is loop 1, the second is loop 2, and so on. The following chart provides a loop
number to controlled process reference for use in selecting the desired parameter from the register
list.
TEMPERATURE/HUMIDITY
MODELS
ALTITUDE MODELS
VTS/TSB
MODELS
DTS MODELS
LOOP
1
TEMPERATURE
TEMPERATURE
TEMPERATURE
TEMPERATURE
H O T C H A M B E R ( B A T H )
LEFT CHAMBER
2
PRODUCT
H U M I D I T Y
ALTITUDE
H U M I D I T Y
COL D C HA MB ER (BA T H)
CENTER CHAMBER
3
-
PRODUCT
PRODUCT
ALTITUDE
P R O D U C T
RIGHT CHAMBER
4
-
-
-
PRODUCT
DUT LEFT BASKET
5
-
-
-
-
DUT RIGHT BASKET
The chamber events also vary based on the model of chamber and options present. In order to turn
the chamber and associated options on and off, it is necessary to set the proper event. The chart
below provides the chamber event number and its associated function based on the chamber model.
The chamber events are listed in order from top to bottom on the EZT-570S “Overview” screen.
STANDARD MODELS
(TMP/RH/ALTITUDE)
VTS/TSB MODELS
DTS MODELS
EVENT
1
CHAMBER
HOT CHAMBER / BATH ON
LEFT CHAMBER
2
HUMIDITY
COLD CHAMBER (VTS ONLY)
CENTER CHAMBER
3
AUX COOL
AUX COOL
AUX COOL
4 PURGE
PURGE
PURGE
5
ALTITUDE
XFR HOT
XFR LEFT
6
-
XFR COLD
XFR RIGHT
7
-
XFR UNLOAD (TSB ONLY)
RIGHT CHAMBER
8 -
-
-
9
INITIATE DEFROST
INITIATE DEFROST (VTS ONLY)
INITIATE DEFROST (CENTER ONLY)
10
PRODUCT CONTROL
PRODUCT CONTROL
PRODUCT CONTROL
11
REMOTE SETPOINT 1
REMOTE SETPOINT 1
REMOTE SETPOINT 1
12
REMOTE SETPOINT 2
REMOTE SETPOINT 2
REMOTE SETPOINT 2
13
REMOTE SETPOINT 3
REMOTE SETPOINT 3
REMOTE SETPOINT 3
14
REMOTE SETPOINT 4
REMOTE SETPOINT 4
REMOTE SETPOINT 4
15
REMOTE SETPOINT 5
REMOTE SETPOINT 5
REMOTE SETPOINT 5