![Cincinnati Sub-Zero EZT-570S Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/cincinnati-sub-zero/ezt-570s/ezt-570s_reference-manual_2607293020.webp)
EZT-570S User Communication Reference Manual
20
Adhere to the following list of registers and their allowable data ranges. Do not attempt to
write to any other register number than those listed below. Do not write to registers that are
for options your chamber does not have. Failure to adhere to this requirement can result in
erratic control and/or damage to equipment.
All register numbers listed are relative values. To convert to absolute values, add 400001.
Register
Address
Parameter Description
Data
*A
Range
*B
*C
Type Low
High Unit
0
(0x0000)
System Mode
R/W
*B1
*B1
-
1
(0x0001)
Clock (Year/Month)
R
*B2
*B2
-
2
(0x0002)
Clock (Day/DOW)
R
*B3
*B3
-
3
(0x0003)
Clock (Hours/Minutes)
R
*B4
*B4
-
4
(0x0004)
Clock (Seconds)
R
0
59
seconds
5
(0x0005)
Power Recovery Mode
R/W
*B5
*B5
-
6
(0x0006)
Power Recovery Time
R/W
0
32767
seconds
7
(0x0007)
Defrost Operating Mode
R/W
*B6
*B6
-
8
(0x0008)
Defrost Temperature Set point
R/W
-3276.8
3276.7
*C1
9
(0x0009)
Defrost Interval
R/W
0
32767
minutes
10
(0x000A)
Defrost Status
R
*B7
*B7
-
11
(0x000B)
Time Remaining Until Next Defrost
R
0
32767
minutes
12
(0x000C)
Product Control
R
*B8
*B8
-
13
(0x000D)
Product Control Upper Set Point
R/W
-3276.8
3276.7
*C1
14
(0x000E)
Product Control Lower Set Point
R/W
-3276.8
3276.7
*C1
15
(0x000F)
Condensation Control (Off/On)
R/W
0
1
-
16
(0x0010)
Condensation Control Monitor Mode
R/W
*B9
*B9
-
17
(0x0011)
Condensation Control Input Selection
R/W
*B10
*B10
-
18
(0x0012)
Condensation Control Ramp Rate Limit
R/W
*B11
*B11
-
19
(0x0013)
Condensation Control Dewpoint Limit
R
-3276.8
3276.7
*C1
20
(0x0014)
Condensation Control Actual Dewpoint R
-3276.8
3276.7
*C1
21
(0x0015)
Chamber Light Control (Off/On)
R/W
0
1
-
22
(0x0016)
Chamber Events
R/W
*B12
*B12
-
23
(0x0017)
Customer Events
R/W
*B12
*B12
-
24
(0x0018)
Program Control/Status
R/W
*B13
*B13
-
25
(0x0019)
Program Advance Previous/Next Step
R/W
*B14
*B14
-
26
(0x001A)
Program Name Characters 1 & 2
R
*B15
*B15
-
27
(0x001B)
Program Name Characters 3 & 4
R
*B15
*B15
-
28
(0x001C)
Program Name Characters 5 & 6
R
*B15
*B15
-
29
(0x001D)
Program Name Characters 7 & 8
R
*B15
*B15
-
30
(0x001E)
Program Name Characters 9 & 10
R
*B15
*B15
-
31
(0x001F)
Year/Month Program Started
R
*B2
*B2
-