Freeze-dryer Delta 1-24 LSCplus
Freeze-dryer Delta 2-24 LSCplus
7 Malfunctions and error correction
Version 04/2013, Rev. 2.9 of 09/03/2020
• sb-dc
87 / 125
Translation of the original operating manual
→
7.1.2
Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is
frozen.
→
7.1.2.1
Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 65: Small flange and centring ring
Fig. 66: Small flange with centring ring
and small flange
Fig. 67: Attaching the clamping ring
Fig. 68: Tightened clamping ring