Digital Blood Pressure Monitor
Digital Blood Pressure Monitor
BP11
BP11
- 14 -
- 15 -
the last 3 readings of the last group used. The “
“ symbol will appear along with
the corresponding WHO Blood Pressure Indicator. The Memory Check mode can
be accessed by pressing the “M “ buttons. To check the average results from other
groups, select the desired group first prior to activating the “ M “ button in the off
position. (See “Select Memory Group “on Page 10.)
Note: In this mode, user can access to test mode by pressing “ M “ button.
Low Battery Indicator
4 short warning beeps sound when battery life is depleting and unable to inflate cuff
for testing. The “
“ appears simultaneously for approximately 5 seconds prior to
shutting off. Replace batteries at this time. No memory loss will occur throughout this
process.
Troubleshooting
Problem
Possible Cause
Solution
Blood pressure
results are not
w i t h i n t y p i c a l
range
Cuff is too tight or not
properly positioned on
the arm
F i r m l y r e p o s i t i o n c u f f
approximately1-2cm1/2" ( 1/2”)
above the elbow joint( See Page 12)
Inaccurate test results
due to body movement
or monitor movement
Sit in a relaxed position with arm
placed near heart.. Avoid speaking
or moving body parts while testing.
Make sure the monitor unit is placed
in a stationary position throughout
the testing period. ( See Page 7)
"
"displayed
C u f f f a i l s t o i n f l a t e
properly
Make sure hose is properly fastened
to cuff and monitor unit
Improper operation
Read user manual carefully and re-
test properly.
Pressurization is over
300mmHg
Read user manual carefully and re-
test properly.