3. Confirm the judgment result.
When the buzzer sounds
“
Be-Beep
”
for the judgment completion, confirm the result by lamps.
Green lamp lighting
Good (The measured value is lower than the set value.)
[Without beep sound]
Red lamp blinking
Failure (The measured value is higher than the set
value.)
[ Be-Be-Beep-Be-Be-Beep
]
Green
/
Red lamps
Both lighting
Measurement error
[ Beeeeep]
It is considered that the measurement error is based on (1) too short distance (less than 10cm), (2) too long
distance (more than 25cm), or (3) movement (position changed) before the judgment completion, etc.
When the measured person leaves from the unit after confirming the judgment result, the unit is reset and
returns to the stand-by.
Since the fever is suspected by blinking of the red lamp, perform the confirmation of the body
temperature by reexamination with a clinical thermometer, etc. This unit is an instrument to measure
the surface temperature of measurement objects for measurement, and is not a clinical thermometer.
■
Connection of the checker and PC and its setting
To change the setting of the unit or confirm the thermal image, connect to the PC and operate by exclusive
application software.
■
LAN connection
Connect this unit and a personal computer with the attached LAN cable.
For the LAN settings, be sure to connect the body surface temperature checker and a personal
computer by 1 to 1. When other instruments (including other body surface temperature checkers) are
connected via a hub etc., be sure to remove them.
Plug
(
100V-240V AC
)
Judgment result
Green
lighting
Red
blinking
The buzzer sounds.
Good
Fever suspected
LAN port
LAN cable
LAN port
Be-Be-Beep-Be-Be-Beep
Measurement
error
(Try again.)
Be-Be-Beep-Be-Be-Beep
Beeeeep
Green
/
red
lighting