FISIOLASER IRD/IRD 2
WARNING
Laser radiation from the equipment is inherently dangerous
:
always use safety goggles;
Do not look at the beam with your naked eye or with any optical instrument
Do not expose the beam unless it is being strictly controlled
avoid exposing the eye to direct or diffuse radiation.
Direct exposure to laser radiation for a fraction of a second is not dangerous,
however prolonged exposure can be harmful.
The damage is directly proportional to the power emitted .
If a channel is in use, you obviously cannot set a new program on it until it finishes.
! WARNING !
To avoid contamination of the use environment for FISIOLASER IRD-series unit
and / or persons involved in its use, do not apply to contact with patients laser
probes that have not been thoroughly cleaned and disinfected at the end of the
previous treatment .
Once the chosen treatment program appears on the display, press the START button
to start it up .
LASER TEST
The laser probe is provided with a contact detector that allows the emission only
when the head of the probe is located in proximity of the treatment point.
It is possible to verify the correct emission, by close the probe to a non-reflecting
surface after setting the device in automatic mode and pushing the start button. The
contact detector once properly detected the surface to treat will automatically
activate the emission of treatment, and the time parameter will start to count.
SETTING UP PROGRAMMES
This function can be only used with the Smart Card inserted as it is the only memory
available for storing new programmes .
For to create a customized program , please follow these instructions :
1)
push the function taste relative to the CREATE PROGRAM button ( see
fig. 4 ) ; it appears an empty list in which you can enter new programs
2)
rotate the encoder knob until to position the cursor on the desired row for to
save the new program (if the smart-card is just partially filled)
3)
press the encoder knob for to confirm the choice ; it appears to video the
next screen (see fig.25) :
Fig. 25
4)
Rotate the encoder knob until to position the cursor on the parameter that
you wants to modify
5)
Press the encoder knob for to confirm the choice to customize the value of
the selected parameter
6)
rotate the encoder knob ( in clockwise way for increasing values , in
counter-clockwise way for decreasing values ) until to reach the desired
value to assign at the selected parameter
7)
press the encoder knob for to confirm the value assignation
8)
repeat the steps from 4) to 7) for all of the parameters that you want to
modify
9)
Rotate the encoder knob until to position the cursor on the last menu of the
page , then press the encoder knob in correspondence of this menu ; it
appears the next screen (see fig.26)
Save
Exit
CREATION PROGRAM