![Chiltern Invadex WISPA 100 Plus Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/chiltern-invadex/wispa-100-plus/wispa-100-plus_user-manual_2601697007.webp)
Page 5
HOIST
100 Series
Issue
2.0
Issue
2.0
June 2018
Emergency Lowering
In the unlikely event of a power failure or hoist malfunction the
hoist occupant may be lowered using the emergency lowering
cord.
To operate, pull the white cord gently until 2 clicks are heard/
felt. The occupant will then be slowly lowered.
In certain circumstances the cord may need to be pulled more
gently for just one click, to lower the occupant. If this occurs
refer to the troubleshooting guide on page 8.
Do not try and lift using the emergency lowering device as this
can damage the hoist.
EMERGENCY
STOP
LOW
ER
Emergency lowering
pull cord
Emergency Stop
To operate in case of an emergency, pull the red cord to switch off the hoist completely.
EMERGENCY
STOP
LOW
ER
Emergency stop
pull cord
RESET
Reset Procedure
Once the emergency stop or overload device have been
activated, the hoist will need to be reset.
To reset slide the plastic tube on the red cord upwards through
the hole in the case and push to reset the switch. A click will
be hear when the switch resets.
The hoist should now operate normally.
Please Note:
The emergency lowering device uses a PP3 battery to allow the occupant
of the hoist to be lowered in the event of an emergency. Frequent testing of this device
in non-emergency situations will quickly discharge the battery. The battery should be
renewed at each service and after use.