5
To use the bowl without selector, firmly grasp the bowl
body and use your other hand to rotate the particle size
regulator (C1.1) clockwise until it is ejected from the up-
per part (C1.2)
C1.1
C1.2
INSTRUCTIONS FOR USE
Before each use, clean hands thoroughly and clean the device as described in the
section on “CLEANING SANITISATION DISINFECTION”. During use, it is advisable to
protect yourself from any dripping. The nebulizer cup and accessories are for personal
use only to avoid potential risks of transmitting infections. This device is suitable for
the administration of medical substances and not, for which the administration via
aerosol is foreseen; these substances are to be in any case prescribed by the Doctor.
With viscous substances, dilution with a suitable physiological solution could be nee-
ded, according to the medical prescription.
1.Plug the power cord (A7) into a power socket cor-
responding to the voltage of the unit. This must
be positioned so that it is not difficult to make the
disconnection from the mains.
2.
Pour the liquid to be nebulised (medical device and/or
the medicine prescribed by the doctor) in the lower part
(C1.4) of the open cup. Insert the upper part (C1.2) of
the cup into the lower part (C1.4), then close the cup by
turning the upper part (C1.2) clockwise.
3. Connect accessories as indicated in the “Connection diagram” on the cover.
4. Set the nebulizer selector to the desired particle size as described under “How to use
the
Clenny
®
A
nebulizer”.
5. Sit comfortably holding the nebulizer in your hand, place the mouthpiece onto your
mouth or alternatively use the nose piece or mask. Should you opt for the mask ac-
cessory, place it on your face as shown in the picture (with or without using the elasti-
cated strap), so that the sides adhere well to the face.
6. Start the unit by means of switch (A1) and Inhale gently and deeply; after inspiration
we recommend to hold your breath for a moment in order to allow inhaled aerosol
droplets to deposit. Then, exhale slowly.
7. When the treatment is finished, turn off and unplug the unit.
Wash your face and
rinse your mouth.
WARNING:
If after the therapy session moisture accumulates in the tube (B), remove
the tube from the nebulizer and dry it using the compressor air flow; this operation
will prevent mold from forming inside the tube.
RESPIRATORY
SYSTEM
UPPER
AIRWAYS
LOWER AIRWAYS
Ear, nose and
throat tract
Trachea and
primary bronchi
Secondary
bronchi
Bronchioles
CONFIGURA-
TION
BOWL
Without
Selector
Selector in
position 1
Selector in
position 2
Selector in
position 3
MMAD
(μm)
6.8
(3)
4.6
(1)
3.9
(1)
2.7
(1)
Output rate
(ml/min’)
0.66
(2)
0.45
(2)
0.41
(2)
0.27
(2)
Breathable
fraction < 5 μm
(FPF)
31.3%
(3)
54.5%
(1)
62.0%
(1)
80.0%
(1)
1) In vitro testing performed by TÜV Rheinland Italia S.r.l. in collaboration with the University of Parma and in
compliance with the EN 13544-1: 2007 + A1 European Standard for aerosol therapy devices Standard. More details
are available on request.
(2) Data measured according to Flaem I29-P07.5 internal procedure Dispensing speed values may vary according
to the patient’s respiratory capacity. More details are available on request.
(3) Data calculated on the basis of the values obtained with Malvern Spraytec, and compared with the values of
the tests performed at the University of Parma.