In most cases if the over temperature protection circuit is active, it must be inactivated before
starting this procedure.
1.
Set the equipment up in the exact configuration it will be used. Ensure all connections to the
Temperature Control Module.
2.
Turn controller on an enter the desired set point temperature. If the set point isn’t at least 30
°
C above
ambient, then skip this procedure and go to the next procedure, ‘Autotuning the Controller for Very
Fine Control’.
To remove the “
tunE
” “
FAiL
” message, turn the controller off for 10 seconds. Try the procedure titled
“Autotuning the Controller for Very Fine Control”. If autotune continues to fail, call and discuss your
application with one of our engineers.