–15 –
5.0 MAKING MEASUREMENTS
The steps involved in making measurements are detailed in this section. The following
sections outline how to setup and prepare your TI-25DLX for field use.
An automatic or manual zero must always be performed. The auto zero is an ‘off block’
electronic zero that does not require a zero reference standard. This will most always be
the zero option of choice, as it makes the zeroing process very easy and convenient to
perform. However, the manual zero option offers better accuracy in terms of a reference
point. If the manual zero option is enabled, the probe zero must be measured on the
reference disk (battery disk) attached to the top of the instrument. The zero compensates
for variations in the transducer. In either mode the sound velocity must be determined,
and is used to convert the transit time to a physical length. The sound velocity can be
selected from a material chart in the manual, selected from a short list of common
materials in the TI-25DLX , or for greater precision determined from a sample of the
test material that has been mechanically measured. To enter the velocity from a table,
look up the material on the chart in the appendix of this manual and refer to the section
below on Calibration to a Known Velocity. To determine the velocity of a single sample,
refer to the Material Calibration section in section 5.2.
When measuring curved materials, it’s more accurate to calibrate from two test points,
one at the minimum limit of the target thickness and one at the maximum limit. In this
case the reference disk mounted to the TI-25DLX is not used. This is called two-point
calibration and is described in section 5.2
5.1 Probe zero
Setting the zero point of the TI-25DLX is important for the same reason that setting
the zero on a mechanical micrometer is important. It must be done prior to calibration,
and should be done throughout the day to account for any temperature changes in the
probe. If the TI-25DLX is not zeroed correctly, all the measurements taken may be in
error by some fixed value. The zero can only be performed with the measurement mode
set to pulse-echo (P-E). Therefore, if the TI-25DLX is to use the echo-echo (E-E)
measurement mode and a manual zero is being performed, the TI-25DLX will argue by
briefly displaying the message “NO PRB0”.
IMPORTANT NOTE:
The internal zero setting of the TI-25DLX , used for the auto
zero mode, can be reset at anytime by performing a “manual zero”, and immediately
followed by performing an “auto zero”.