1.0 GENERAL
The pins of the CP-425-KIT have to be free from dust, dirt, oil and dampness.
The ideal test temperature was defined at 23°C. However, a test temperature
between 20°C and 26°C is sufficient. Air humidity has to be between 30% r.h. and
80% r.h.
The device as well as the CP-425-KIT has to be stored in the same surroundings for
some time before testing. If you do not comply with these specifications this may
lead to deviations.
This manual declares the test of following devices:
• TMT-425 Digital Textile Moisture Meter
• TMT-425-USB Digital Textile Moisture Meter
2.0
TEST PROCEDURE
Switch on the TMT-425(USB) and switch to the calibration curve „test block“ using
the arrow keys.
Put the
orange stick
(CP-425-1) into the electrode holder.
free from dust, dirt, oil and dampness