![Chauvin Arnoux F 1N Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/chauvin-arnoux/f-1n/f-1n_user-manual_2594606050.webp)
50
Specifications in normal use (except during RECORD).
RANGE
40.00 A
400.0 A
WORKING
50 mA
100 mA
RANGE
100 A
700 A
SPECIFIED MEASURE-
300 mA
100 A
400 A
MENT RANGE
99.99 A
400 A
700 A
MEASUREMENT ERROR
2 % rdg + 20 pts
2 % rdg
5 % rdg
rdg : reading
rdg : reading
Remark: this error is given (±).
The current detection threshold is 50 mA.
■
Stability of digital display (up to 4000 counts): 2 counts
Measurement resolution on digital display: 1 count
■
Display response time:
analog: from 100 to 150 ms to 95 %.
digital: 500 ms to 90 % / 2 s to 99 %.
Specifications on "RECORD 100 ms" mode (MAX, MIN, AVG and present value)
The specifications of the present value and the average value (AVG) are identical to those
in normal use.
Specifications of MAX and MIN values:
■
Accuracy: identical to normal mode.
■
Stability, repeatability: 10 counts.
■
MAX/MIN measurement (for a signal in the measurement range):
200 ms to 300 ms for currents above 8 A peak approx.
1 to 2 s for currents under 8 A peak (MAX and MIN taken on the current value).
Specifications in PEAK and RECORD PEAK
Resolution of digital display:
20 counts up to 4000 counts.
40 counts from 4000 to 10 000 counts.
RANGE
40.00 A
400.0 A
WORKING
300 mA
300 mA
RANGE
100 A
1000 A
SPECIFIED MEASURE-
400 mA
100 A
600 A
MENT RANGE
99.99 A
600 A
1000 A
MEASUREMENT ERROR
3 % rdg + 30 pts
3 % rdg
8 % rdg