Bargraph
Specified measurement range
0,1 M
Ω
- 50 G
Ω
*
Resolution
9 segments per decade
Intrinsic uncertainty
± (5% + 1 segment)
**: When the measurement range is exceeded, the whole bargraph is displayed.
Test voltage
With a test current < 1mA, the intrinsic uncertainty on U
N
is ± 0,5V.
Specified measurement range
0.0 - 100.0 V
Resolution
0,1 V
Intrinsic uncertainty
± (3% + 3 ct)
Typical discharge time after test
To go from U
N
to 25 V, the discharge time is < 2s/µF
Test current
Maximum test current: 2mA +0% -50%
Specified measurement range
0.01 - 39.99 µA
40.0 - 399.9 µA
0.400 - 2.000 mA
Resolution
10 nA
100 nA
1 µA
Intrinsic uncertainty
± (10% + 3 ct)
Typical test voltage vs load curve
The voltage as a function of the measured resistance takes the following form:
I = 1 mA
U
N
R
N
= U
N
/ 1 mA
The range of operation per IEC 61557 is from 100k
Ω
to 2 G
Ω
(see § 3.4).
3.2.3. CONTINUITY MEASUREMENTS
Particular reference conditions
Inductance in series with the resistance: zero.
Specified measurement range
(without compensation of the leads)
0.00 * - 10.00
Ω
0.0 * - 100.0
Ω
Resolution
10 m
Ω
100 m
Ω
Intrinsic uncertainty
± (2% + 2 ct)
Test current
200 mA
20 mA
No-load voltage
≥
6 V