15
0
> 2s
M
mA
The display changes to zero and the
symbol is displayed. The resistance of the leads will be systematically subtracted from
all continuity measurements. If the resistance of the leads is > 10
Ω
, there is no compensation.
The compensation remains in memory until the instrument is switched off. The continuity measurement range is reduced
by the stored compensation value.
If the leads are changed with no change of compensation, the display may become negative. The instrument reports that
the compensation must be redone by displaying
blinking.
2.4.2. ELIMINATION OF THE COMPENSATION OF THE LEADS
To eliminate the compensation of the leads, leave the leads open and long-press the
0
key.
0
> 2s
2.4.1. COMPENSATION OF THE LEADS
To ensure precise measurements, it is necessary to compensate the resistance of the measurement leads.
Short-circuit the measurement leads and long-press the
0
key.
The display indicates the resistance of the leads and the
symbol goes off.