3. TECHNICAL CHARACTERISTICS
3.1. GENERAL REFERENCE CONDITIONS
Quantity of influence
Reference values
Temperature
23 ± 3 °C
Relative humidity
45 to 55% RH
Frequency
DC and 45 to 65 Hz
Supply voltage
8 ± 0.2V
battery life indication 58 ± 8%
Electric field
0V/m
Magnetic field
< 40A/m
The intrinsic uncertainty
is the error specified for the reference conditions.
The operating uncertainty
includes the intrinsic uncertainty plus variations of the quantities of influence (position, supply voltage,
temperature, etc.) as defined in standard IEC-61557.
The uncertainties are expressed in % of the reading (R) and in number of display points (ct):
± (a %R + b ct)
3.2. ELECTRICAL CHARACTERISTICS
3.2.1. VOLTAGE MEASUREMENTS
Particular reference conditions
Peak factor = 1.414 in AC, sinusoidal signal
Specified measurement range
0.3 - 399.9V
400 - 700V
Resolution
0.1V
1V
Intrinsic uncertainty
± (3% + 2 ct)
Input impedance
400k
Ω
Frequency ranges
DC and 15.3 at 800Hz
3.2.2. INSULATION MEASUREMENT
Particular reference conditions
Capacitance in parallel on resistance: null
Measurement range
Test voltage (U
N
)
R
10 V - 100 V
(U
N
/5) k
Ω
- (U
N
/5) G
Ω
Intrinsic uncertainty
Test voltage (U
N
)
10V ... 100V
Specified measurement
range
2 - 999 k
Ω
et 1.000
- 3.999 M
Ω
4.00 - 39.99 M
Ω
40.0 - 399.9 M
Ω
400 - 3999 M
Ω
4.00 - 20.00 G
Ω
Resolution
1 k
Ω
10 k
Ω
100 k
Ω
1 M
Ω
10 M
Ω
Intrinsic uncertainty
± (3% + 2 ct + (10%/U
N
) per 100 M
Ω
)
With a test voltage
≥
50 V and an insulation resistance
≤
2 G
Ω
, the intrinsic uncertainty is ± (3% + 2 ct).