30
The range of operation per IEC 61557 is from 100k
:
to 2 G
:
(see § 4.2).
3.2.4. CONTINUITY MEASUREMENTS
Particular reference conditions
Inductance in series with the resistance: zero.
Specified measurement range (without
compensation of the leads)
0.00 * - 10.00
:
0.0 * - 100.0
:
Resolution
10 m
:
100 m
:
Intrinsic uncertainty
± (2% + 2 ct)
Test current
200 mA
20 mA
No-load voltage
t
6 V
*: In the case of incorrect compensation of the leads, the instrument allows display of negative values, down to -0.05
:
at 200 A
and -0.5
:
at 20 mA.
Test current
200 mA range: 200mA (-0mA + 20mA)
20 mA range: 20mA ± 5mA
Specified measurement range
0 - 250 mA
Resolution
1 mA
Intrinsic uncertainty
± (2 % + 2 ct)
Compensation of the leads:
0 to 9.99
:
.
3.2.5. RESISTANCE MEASUREMENTS
Specified measurement range
0 - 3999
:
4.00 - 39.99 k
:
40.0 - 399.9 k
:
400 - 1000 k
:
Resolution
1
:
10
:
100
:
1 k
:
Intrinsic uncertainty
± (3% + 2 ct)
No-load voltage
approximately 4,5 V
3.2.6. CAPACITANCE MEASUREMENTS (C.A 6532)
Capacitance
Specified measurement range
0.1 - 399.9 nF
400 - 3999 nF
4.00 - 10.0 µF
Resolution
0.1 nF
1 nF
10 nF
Intrinsic uncertainty
± (3% + 2 ct)
Line length
Capacitance per unit length: 40 to 60 nF/km (50 nF/km is default)
Specified measurement range
0.000 - 3.999 km
4.00 - 39.99 km
40.0 - 100.0 km
Resolution
1 m
10 m
100 m
Intrinsic uncertainty
± (3% + 2 ct)