8. tECHnICAl CHARACtERIstICs
8.1. GEnERAl REfEREnCE ConDItIons
Quantity of influence
Reference values
Temperature
20 ± 3 °C
Relative humidity
45 to 55 % HR
Supply voltage
9.6 ± 0.2 V
Frequency
DC and 45 to 65 Hz
Electric field
< 1 V/m
Magnetic field
< 40 A/m
Supply
on battery (mains not connected)
the intrinsic uncertainty
is the error defined under the reference conditions.
the operating uncertainty
includes the intrinsic uncertainty plus the effects of variation of the quantities of influence (supply
voltage, temperature, interference, etc.) as defined in standard IEC-61557.
8.2. ElECtRICAl CHARACtERIstICs
8.2.1. voltAGE MEAsUREMEnts
Particular reference conditions:
Peak factor = 1.414 in AC (sinusoidal signal)
AC component <0.1% in DC measurement
DC component <0.1% in AC measurement
Measurement range (AC or DC)
0.2 - 399.9 V
2.0 - 399.9 V
400 - 550 V
Resolution
0.1 V
1 V
Intrinsic uncertainty
± (1.5 % + 2 ct)
± (1.5 % + 1 ct)
Input impedance
450 k
W
Frequency of use
DC and 15.8 … 450 Hz
Contact voltage measurements
Measurement range (AC)
2.0 - 100.0 V
Intrinsic uncertainty
± (15% + 2 ct)
Input impedance
6 M
W
Frequency of use
15.8 … 65 Hz
This voltage is displayed only if it exceeds U
L
(25 V, 50 V or 65 V).
Measurements of potential of the voltage probe
The characteristics are the same as in the voltage measurements.
This voltage must normally be between 0 and U
L
.