SAFETy iNSTruCTiONS
10
Before connecting the unit, make sure that:
• The voltage rating on the safety label corresponds to
the available system voltage.
• The frequency rating on the rating plate corresponds to
the system frequency.
• An earthed socket outlet with earthing contact is
available for connecting the unit.
• The routing of the power cable from the unit to the
socket outlet with earthing contact does not pose a
danger for personnel or the patient.
• The Mains Power Supply is designed for the
comparatively high (possibly additional) power input of
the unit (~ 1500VA) and the line is sufficiently protected
in accordance with regulations.
PrOTECTiON OF ThE uNiT
Improper installation, operation or maintenance of the
shortwave therapy unit may result in malfunctions of this
unit or other devices.
Observe the following instructions in order to prevent
malfunctions:
•
In order to prevent electromagnetic disturbances,
place the unit at least 6 meters from any other devices.
Also make sure that there is sufficient distance
between the unit and Mains Power Supply or data
cables in walls, ceilings and floors, since the
electromagnetic radiation from the unit can pass these
essentially without hindrance.
•
In selecting the location for the unit, make sure that
the patient has contact during the treatment to the
non-earthed application element and, due to
equalizing currents in case of differing potentials, that
the patient is never in contact with metal elements
(especially if they are earthed), such as heating
radiators, metal beds or other earthed devices.
Make certain that the unit is electrically earthed
by connecting only to a earthed electrical service
receptacle, conforming to the applicable national and
local electrical codes.