Intelect® Mobile Ultrasound
ABOUT ULTRASOUND THERAPY
DANGER
DANGER
•
Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air, with oxygen or nitrous
oxide.
•
The solvents of adhesives and flammable solutions used for
cleaning and disinfecting should be allowed to evaporate
before the unit is used.
•
DO NOT connect the unit to an electrical supply without
first verifying that the power supply is the correct voltage.
Incorrect voltage may cause unit damage, malfunction,
electrical shock, fire, or personal injury. Your unit was
constructed to operate only on the electrical voltage
specified on the Voltage Rating and Serial Number Plate.
Contact your DJO dealer if the unit is not properly rated.
•
NiMH Batteries contain Class E Corrosive materials. In the
event of battery cell rupture or leakage, handle battery pack
wearing neoprene or natural rubber gloves. Contents of a
ruptured or leaking battery can cause respiratory irritation.
Hypersensitivity to nickel can cause allergic pulmonary
asthma. Contents of cell coming in contact with skin can
cause skin irritation and/or chemical burns.
•
Never, under any circumstances, open the battery pack
housing or cells. Should an individual battery from a battery
pack become disassembled, spontaneous combustion of the
negative electrode is possible. There can be a delay between
exposure to air and spontaneous combustion.
•
Charge the battery pack according to the instructions found
in this manual. Never attempt to charge the battery pack on
any other charging mechanism.
•
Use the battery pack only with the Intelect Mobile Series units.
•
Do not reverse the polarity of the battery pack. Doing so can increase the
individual cell temperature and cause cell rupture or leakage.
•
Never dispose of the battery pack in fire. Never short circuit the battery
pack. The battery pack may explode, ignite, leak or get hot causing
serious personal injury.
•
Dispose of NiMH batteries according to national, state and local codes
and regulations.