41
Intelect® Mobile Ultrasound
The Operator Interface consists of buttons with a liquid crystal display (LCD) . The operator is able to view parameter options on the
display and make selections by pressing the buttons on the control panel . The LCD will provide continuous information during the
treatments concerning power and elapsed time . Parameters are adjusted using control panel buttons on the front of the unit . The
ultrasound output can be stopped by pressing the “PAUSE” or “STOP” buttons located on the control panel .
To apply ultrasound therapy, do the following:
1 .
Turn system power “ON" by pressing the Power On/Off Button .
The
unit will go through self diagnostics, and the home screen displays on the LCD .
2 .
Press the Frequency button to select 1 or 3 MHz . When the button is pressed, the
frequency will toggle from 1 to 3 MHz and back again as long as the button is being
pressed .
An audible tone will be heard when changes are made .
NOTE: With 1 cm
2
, 2 cm
2
, and 5 cm
2
sound head, switching from 1 to 3 MHz results in
no change in power (only 3 .3 MHz is available with the 1 cm
2
sound head) . When using
a 10 cm
2
head with greater than 10 watts, changing from 1 to 3 MHz reduces power to
10
watts .
STARTINg, STOPPINg, AND INTERRUPTINg THERAPY
OPERATION