Intelect® Mobile Ultrasound
CAUTION
•
DO NOT disassemble, modify, or remodel the unit or accessories. This may
cause unit damage, malfunction, electrical shock, fire, or personal injury.
•
Failure to use and maintain the Intelect Mobile Ultrasound and its
accessories in accordance with the instructions outlined in this manual
will invalidate your warranty.
•
DO NOT remove the cover. This may cause unit damage, malfunction,
electrical shock, fire, or personal injury. There are no user-serviceable parts
inside the unit. If a malfunction occurs, discontinue use immediately and
consult the dealer for repair service.
•
DO NOT permit any foreign materials or liquids to enter the unit. Take
care to prevent any foreign materials including, but not limited to,
inflammables, water, and metallic objects from entering the unit. These
may cause unit damage, malfunction, electrical shock, fire, or personal
injury.
•
If you have difficulty operating the unit after carefully reviewing this
operator’s guide, contact your DJO dealer for assistance.
•
This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instructions, may
cause harmful interference to other devices in the vicinity. However, there
is no guarantee that interference will not occur in a particular installation.
Harmful interference to other devices can be determined by turning
this equipment on and off, try to correct the interference using one or
more of the following: Reorient or relocate the receiving device, increase
the separation between the equipment, connect the equipment to an
outlet on a different circuit from that to which the other device(s) are
connected and/or consult the factory field service technician for help.
•
Use of parts or materials other than DJO's can degrade minimum safety.
•
Always keep the sound head in constant motion.
•
Always keep the sound head in full contact with the patient’s skin or
submerged under water when setting intensity.
•
Use ample conductive gel to ensure good coupling throughout the
treatment. If needed, apply when setting intensity.
•
Be sure to read all instructions for operation before treating a patient.
•
Do not drop the applicator on hard surfaces. Do not cool an overheated
sound head with ice water or ice packs. Do not allow the sound head
to reach maximum temperatures repeatedly. All of these conditions are
likely to damage the sound head crystal. Damage resulting from these
conditions is not covered under the warranty.
•
Use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous exposure to ultrasonic
energy.
CAUTION
•
Do not operate this unit in an environment where other devices are being
used that intentionally radiate electromagnetic energy in an unshielded
manner. Medical Electrical Equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the EMC
information provided in this manual.
•
Remove battery pack if unit is not to be used for an extended period.
WARNING