Intelect® Mobile Stim
SAFETy PrECAuTiONS
6
WARNING
WARNING
•
Stimulation should not be applied over swollen, infected, and
inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis,
varicose veins, etc.
•
Stimulation should not be applied over, or in proximity to, cancerous
lesions.
•
Output current density is inversely related to electrode size.
Improper application may result in patient injury. If any question
arises as to the proper electrode size, consult a licensed practitioner
prior to therapy session.
•
Dispose of all products in accordance with local and national
regulations and codes.
•
For continued protection against fire hazard, charge the battery
pack only while installed on the Intelect Mobile Stim.
•
Use of controls or adjustments or performance of procedures other
than those specified herein may result in hazardous conditions
causing damage to the battery pack or cells.
•
To prevent electrical shock, disconnect the battery pack from the
system before attempting any maintenance procedures.
•
This equipment is not designed to prevent the ingress of water
or liquids. Ingress of water or liquids could cause malfunction of
internal components of the system and therefore create a risk of
injury to the patient.
wArNiNgS (continued)
•
Use only accessories that are specially designed for this device.
Do not use accessories manufactured by other companies on this
device. DJO, LLC is not responsible for any consequence resulting
from using products manufactured by other companies. The use
of other accessories or cables may result in increased emissions or
decreased immunity of this device.
•
If unit is not in use, power off unit or remove Lead Wires.
•
When the unit is not in use, it should be protected against
unqualified use.
•
Keep electrodes separated during treatment. Electrodes in contact
with each other could result in improper stimulation or skin burns.
•
Output current density is inversely related to electrode size (i.e.,
the larger the electrode, the lower the current density). Improper
application may result in patient injury. If any question arises as to
the proper electrode size, consult a licensed practitioner prior to
therapy session.
•
Powered muscle stimulators should be used only with the leads
and electrodes recommended for use by the manufacturer.
•
Do not turn the unit on or off while it is connected to the patient.